Doppler haemodynamic profiles of clinically and echocardiographically normal mitral and aortic valve prostheses
The Department of Medicine, Clinical Cardiology and the Department of Surgery, Cardiothoracic Unit, Hammersmith Hospital R.P.M.S., London W120NN, U.K.
Received 12 December 1990; revised 11 June 1991; .
Correspondence. Petroa Nihoyannopoulos, MD, Department of Medicine, Clinical Cardiology, Hammersmith Hospital. R.P.M.S., London W120NN, U.K.
Abstract
Doppler echocardiographic studies were performed in 380 consecutive patients with 415 normally functioning artificial valves to establish normal Doppler characteristics for each type of prostheses used in our institution, with particular reference to Starr-Edwards valves, and to serve as control studies for future assessment. None of the patients were in heart failure at the time of the study. Peak transaortic velocities (m.s1) were higher and effective orifice areas (cm2) smaller in mechanical valves as a whole, when compared with bioprostheses (P < 0.01); 2.7 ± 0.7 and 1.4 ± 0.55 for Starr-Edwards, 2.7±0.6 and 1.5±0.6 for Bjork-Shiley, 1.8±0.1 and 1.5±0.6 for Duromedics and 1.5±0.06 and 2±0.12 for bioprostheses, respectively. In the mitralposition, the average peak diastolic velocities (m.s1) and pressure halftimes (ms) were higher in mechanical valves, but there was a large overlap between the various types and sizes of prostheses (P = NS), 1.6 ± 0.3 and 98 ± 25 for Starr-Edwards, 1.4 ± 0.3 and 88 ± 26 for Bjork-Shiley, 1.8 ± 0.1 and 75 ± 5 for Duromedics and 1.5 ± 0.3 and 90 ± 20 for bioprostheses, respectively. There was an inverse relation between valve size and pressure halftime for Starr-Edwards prostheses (P <0.01). Doppler flow characteristics in mechanical valves where similar in patients with normal and dysfunctioning prostheses. Valvular or myocardial dysfunction could best be ascertained when early postoperative studies were a vailable for comparison.
No single Doppler measurement was able to separate normal from abnormal prosthetic valve function, so it is recommended that patients have a comprehensive baseline cross-sectional and Doppler examination before postoperative discharge so that each will have a personal Doppler identity. This will help the recognition of possible future prosthetic valve or myocardial deterioration.
Key Words: Prosthetic valves Doppler echocardiography
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
I Ikonomidis, A Tsoukas, F Parthenakis, A Gournizakis, A Kassimatis, L Rallidis, and P Nihoyannopoulos Four year follow up of aortic valve replacement for isolated aortic stenosis: a link between reduction in pressure overload, regression of left ventricular hypertrophy, and diastolic function Heart, September 1, 2001; 86(3): 309 - 316. [Abstract] [Full Text] [PDF] |
||||
![]() |