The accuracy of decision-making of a semi-automatic defibrillator during cardiac arrest
Regional Medical Cardiology Centre, Royal Victoria Hospital Belfast BT12 6BA, N. Ireland
*Department of Bioengineering, University of Ulster Jordanstown, N. Ireland
Received 4 January 1991; revised 22 May 1991; .
Correspondence: Dr A. A. J. Adgey, Regional Medical Cardiology Centre, Royal Victoria Hospital, Belfast BT12 6BA, N. Ireland.
Abstract
The accuracy of decision-making of a semi-automatic defibrillator was assessed at 57 cardiac arrests in 55 patients. The initial rhythm was ventricular fibrillation at 40 arrests and was correctly identified in 37 (sensitivity 92.5%). In theother 17 arrests initial rhythms were not ventricular fibrillation and 16 were correctly identified (specificity 94%). Continuous electrocardiographic recordings of the arrests were analysed in 812 s segments: of 432 segments showing ventricular fibrillation, 352 were correctly identified (sensitivity 81%); 3895 of 4139 segments showing rhythms other than ventricular fibrillation were correctly identified (specificity 94%). The percentage accuracy of detection of ventricular fibrillation varied from 0100% (mean 91%) and for non-ventricular fibrillation from 51100% (mean 92%). When 104 segments of ventricular fibrillation and 470 segments of non-ventricular fibrillation with cardiopulmonary resuscitation artefact were excluded from the analysis, a sensitivity of 91% and specificity of 96% were obtained. Thus the detection of cardiac arrest rhythms using a microprocessor based detection system for patients with cardiac arrest has a high sensitivity for ventricular fibrillation and specificity for non-ventricular fibrillation.
Key Words: Semi-automatic defibrillator ventricular fibrillation non-ventricular fibrillation cardiac arrests
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
W. H Ibrahim Recent advances and controversies in adult cardiopulmonary resuscitation Postgrad. Med. J., October 1, 2007; 83(984): 649 - 654. [Abstract] [Full Text] [PDF] |
||||
![]() |
Part 5: Electrical Therapies: Automated External Defibrillators, Defibrillation, Cardioversion, and Pacing Circulation, December 13, 2005; 112(24_suppl): IV-35 - IV-46. [Full Text] [PDF] |
||||
![]() |
F. Cecchin, D. B. Jorgenson, C. I. Berul, J. C. Perry, A. A. Zimmerman, B. W. Duncan, F. M. Lupinetti, D. Snyder, T. D. Lyster, G. L. Rosenthal, et al. Is Arrhythmia Detection by Automatic External Defibrillator Accurate for Children? : Sensitivity and Specificity of an Automatic External Defibrillator Algorithm in 696 Pediatric Arrhythmias Circulation, May 22, 2001; 103(20): 2483 - 2488. [Abstract] [Full Text] [PDF] |
||||