Experience with pectoral versus abdominal implantation of a small defibrillator
A multicenter comparison in 778 patients
accepted January 25, 1998
Aims
The aim of the study was to examine the results of implanting small sized cardioverter defibrillators in the pectoral as opposed to the abdominal area. Hitherto, owing to the large size of the early defibrillators, the site of implantation had been confined to the abdomen.
Methods
Between 30 March 1993 and 1 November 1994, 778 patients from 63 centres in 14 countries underwent their first device implantation. The study was set up to evaluate the safety and the efficacy of Medtronic models 7219D, a multi-lead abdominal/pectoral implantable cardioverter defibrillator, and 7219C, a pectoral single-lead Active CanTMimplantable cardioverter defibrillator. There were 155 abdominal and 623 pectoral implants. Survival data were comparable during a mean follow-up period of 4·0±4·6 months, with no difference regarding the pectoral placement of single (n=392) or multi-lead (n=231) devices. The only significant difference was related to severe lead-related events: 5·3% in the pectoral vs 11·6% in the abdominal group (P<0·05). These events were mainly related to lead dislodgement. KaplanMeier estimates showed that both single and multi-lead systems, in either the pectoral or abdominal position, demonstrated a similar severe adverse event-free survival.
Conclusion
These findings suggest that an implantable cardioverter defibrillator (18mm thick, 80cc volume, 129g weight) can be implanted in the pectoral position without an increase in clinically relevant adverse events compared to abdominal implantation. Pectoral implantation was associated with a significantly reduced lead-related severe adverse event rate.
Key Words: Implantable cardioverter defibrillator pectoral implantation abdominal implantation ventricular tachyarrhythmia
f1 Correspondence: Ellen Hoffmann, MD, Department of Medicine I, Grosshadern Hospital, Ludwig-Maximilians-University, Marchioninistr. 15, 81377 Munich, Germany.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
R. Becker, J. Ruf-Richter, J. C. Senges-Becker, A. Bauer, S. Weretka, F. Voss, H. A. Katus, and W. Schoels Patient alert in implantable cardioverter defibrillators: toy or tool? J. Am. Coll. Cardiol., July 7, 2004; 44(1): 95 - 98. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Rosenqvist, T. Beyer, M. Block, K. den Dulk, J. Minten, and F. Lindemans Adverse Events With Transvenous Implantable Cardioverter-Defibrillators : A Prospective Multicenter Study Circulation, August 18, 1998; 98(7): 663 - 670. [Abstract] [Full Text] [PDF] |
||||