Skip Navigation

European Heart Journal 1998 19(7):1085-1098; doi:10.1053/euhj.1998.1060
Copyright © 1998 by the European Society of Cardiology.
This Article
Right arrow FREE Full Text (PDF) Freely available
Right arrow E-letters: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when E-letters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (9)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Hoffmann, E
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hoffmann, E
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

Experience with pectoral versus abdominal implantation of a small defibrillator

A multicenter comparison in 778 patients

E Hoffmannf1, G Steinbeck and European Jewel Investigators

accepted January 25, 1998

Aims

The aim of the study was to examine the results of implanting small sized cardioverter defibrillators in the pectoral as opposed to the abdominal area. Hitherto, owing to the large size of the early defibrillators, the site of implantation had been confined to the abdomen.

Methods

Between 30 March 1993 and 1 November 1994, 778 patients from 63 centres in 14 countries underwent their first device implantation. The study was set up to evaluate the safety and the efficacy of Medtronic models 7219D, a multi-lead abdominal/pectoral implantable cardioverter defibrillator, and 7219C, a pectoral single-lead Active CanTMimplantable cardioverter defibrillator. There were 155 abdominal and 623 pectoral implants. Survival data were comparable during a mean follow-up period of 4·0±4·6 months, with no difference regarding the pectoral placement of single (n=392) or multi-lead (n=231) devices. The only significant difference was related to severe lead-related events: 5·3% in the pectoral vs 11·6% in the abdominal group (P<0·05). These events were mainly related to lead dislodgement. Kaplan–Meier estimates showed that both single and multi-lead systems, in either the pectoral or abdominal position, demonstrated a similar severe adverse event-free survival.

Conclusion

These findings suggest that an implantable cardioverter defibrillator (18mm thick, 80cc volume, 129g weight) can be implanted in the pectoral position without an increase in clinically relevant adverse events compared to abdominal implantation. Pectoral implantation was associated with a significantly reduced lead-related severe adverse event rate.

Key Words: Implantable cardioverter defibrillator • pectoral implantation • abdominal implantation • ventricular tachyarrhythmia

f1 Correspondence: Ellen Hoffmann, MD, Department of Medicine I, Grosshadern Hospital, Ludwig-Maximilians-University, Marchioninistr. 15, 81377 Munich, Germany.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
J Am Coll CardiolHome page
R. Becker, J. Ruf-Richter, J. C. Senges-Becker, A. Bauer, S. Weretka, F. Voss, H. A. Katus, and W. Schoels
Patient alert in implantable cardioverter defibrillators: toy or tool?
J. Am. Coll. Cardiol., July 7, 2004; 44(1): 95 - 98.
[Abstract] [Full Text] [PDF]


Home page
CirculationHome page
M. Rosenqvist, T. Beyer, M. Block, K. den Dulk, J. Minten, and F. Lindemans
Adverse Events With Transvenous Implantable Cardioverter-Defibrillators : A Prospective Multicenter Study
Circulation, August 18, 1998; 98(7): 663 - 670.
[Abstract] [Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.