Skip Navigation

European Heart Journal 2001 22(15):1337-1342; doi:10.1053/euhj.2000.2516
Copyright © 2001 by the European Society of Cardiology.
This Article
Right arrow FREE Full Text (PDF) Freely available
Right arrow References
Right arrow E-letters: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when E-letters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (9)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Chiladakis, J.A
Right arrow Articles by Manolis, A.S
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Chiladakis, J.A
Right arrow Articles by Manolis, A.S
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

In-vivo testing of digital cellular telephones in patients with implantable cardioverter-defibrillators

J.A Chiladakis, P Davlouros, G Agelopoulos and A.S Manolisf1

Cardiology Division, Patras University Medical School, Rio, Patras, Greece

revised October 24, 2000; accepted November 1, 2000

Abstract

Aims To investigate the susceptibility of implantable cardioverter defibrillators to electromagnetic interference generated by digital cellular telephones, functioning in both international transmission technologies: the Global System for Mobile Communication (GSM) and the Digital Cellular System (DCS 1800).

Methods and Results In 36 patients with transvenous implantable cardioverter defibrillators from two manufacturers (Medtronic and Guidant/CPI), cellular telephones with different levels of minimal and maximal power output were tested in the transmitting and receiving mode. Evaluation was performed in activated implantable defibrillators during spontaneous cardiac activity and continuous VVI or DDD pacing to assess possible electromagnetic interference. In two patients, appropriateness of ventricular fibrillation detection and therapy was judged during telephone testing. There was no damage, reprogramming, inappropriate shock therapy or pacing inhibition during the tests. In seven pre-pectoral Medtronic implantable defibrillators, transient electromagnetic interference caused 19 erroneous sensing events, when the operating phone was held in close vicinity to the programmer head. These ‘pseudo-oversensing’ events, which did not result in logging of arrhythmia episodes in the device counter, were interpreted as an adverse interaction between the telephone and the programming device.

Conclusion Digital cellular telephones do not represent a risk to Medtronic and Guidant/CPI recipients of the specific implantable defibrillator models herein tested.

Key Words: Implantable cardioverter-defibrillators, electromagnetic interference, digital cellular telephones

f1 Correspondence: Antonis S. Manolis, MD, 41 Kourempana Street, Agios Dimitrios, Athens 17343, Greece.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
J Am Coll CardiolHome page
C. Kolb, S. Schmieder, G.u. Lehmann, B. Zrenner, M. R. Karch, A. Plewan, and C. Schmitt
Do airport metal detectors interfere with implantable pacemakers or cardioverter-defibrillators?
J. Am. Coll. Cardiol., June 4, 2003; 41(11): 2054 - 2059.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart JHome page
M. Niehaus, R. Gille, R. Cierpka, T. Korte, and J. Tebbenjohans
Interference of two common European digital cellular phones with implantable cardioverter-defibrillators
Eur. Heart J., April 1, 2002; 23(7): 586 - 588.
[Full Text] [PDF]


Home page
Eur Heart JHome page
A.S. Manolis and J.A. Chiladakis
A reply
Eur. Heart J., April 1, 2002; 23(7): 588 - 588.
[Full Text] [PDF]


Home page
Eur Heart JHome page
M. Santini
Digital cellular telephones and ICDs
Eur. Heart J., August 1, 2001; 22(15): 1251 - 1252.
[Abstract] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.