In-vivo testing of digital cellular telephones in patients with implantable cardioverter-defibrillators
Cardiology Division, Patras University Medical School, Rio, Patras, Greece
revised October 24, 2000; accepted November 1, 2000
Abstract
Aims To investigate the susceptibility of implantable cardioverter defibrillators to electromagnetic interference generated by digital cellular telephones, functioning in both international transmission technologies: the Global System for Mobile Communication (GSM) and the Digital Cellular System (DCS 1800).
Methods and Results In 36 patients with transvenous implantable cardioverter defibrillators from two manufacturers (Medtronic and Guidant/CPI), cellular telephones with different levels of minimal and maximal power output were tested in the transmitting and receiving mode. Evaluation was performed in activated implantable defibrillators during spontaneous cardiac activity and continuous VVI or DDD pacing to assess possible electromagnetic interference. In two patients, appropriateness of ventricular fibrillation detection and therapy was judged during telephone testing. There was no damage, reprogramming, inappropriate shock therapy or pacing inhibition during the tests. In seven pre-pectoral Medtronic implantable defibrillators, transient electromagnetic interference caused 19 erroneous sensing events, when the operating phone was held in close vicinity to the programmer head. These pseudo-oversensing events, which did not result in logging of arrhythmia episodes in the device counter, were interpreted as an adverse interaction between the telephone and the programming device.
Conclusion Digital cellular telephones do not represent a risk to Medtronic and Guidant/CPI recipients of the specific implantable defibrillator models herein tested.
Key Words: Implantable cardioverter-defibrillators, electromagnetic interference, digital cellular telephones
f1 Correspondence: Antonis S. Manolis, MD, 41 Kourempana Street, Agios Dimitrios, Athens 17343, Greece.
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
C. Kolb, S. Schmieder, G.u. Lehmann, B. Zrenner, M. R. Karch, A. Plewan, and C. Schmitt Do airport metal detectors interfere with implantable pacemakers or cardioverter-defibrillators? J. Am. Coll. Cardiol., June 4, 2003; 41(11): 2054 - 2059. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Niehaus, R. Gille, R. Cierpka, T. Korte, and J. Tebbenjohans Interference of two common European digital cellular phones with implantable cardioverter-defibrillators Eur. Heart J., April 1, 2002; 23(7): 586 - 588. [Full Text] [PDF] |
||||
![]() |
A.S. Manolis and J.A. Chiladakis A reply Eur. Heart J., April 1, 2002; 23(7): 588 - 588. [Full Text] [PDF] |
||||
![]() |
M. Santini Digital cellular telephones and ICDs Eur. Heart J., August 1, 2001; 22(15): 1251 - 1252. [Abstract] [PDF] |
||||