European Heart Journal Advance Access originally published online on November 11, 2008
European Heart Journal 2009 30(8):940-949; doi:10.1093/eurheartj/ehn481
Toward understanding response to cardiac resynchronization therapy: left ventricular dyssynchrony is only one of multiple mechanisms
1 Department of Cardiology, St Georges Hospital, Blackshaw Road, SW17 0QT London, UK
2 ICREA, Universitat Pompeu Fabra (CISTIB), Barcelona, Spain
3 University of Leuven, Belgium
4 Medical Centre Rijnmond South, Rotterdam, The Netherlands
5 Hopital Pontchaillou, Rennes, France
6 Hopital Louis Pradel, Lyon, France
Received 16 April 2008; revised 2 September 2008; accepted 2 October 2008; online publish-ahead-of-print 11 November 2008.
* Corresponding author. Tel: +44 2087251397, Fax: +44 2087254402, Email: chirineparsai{at}hotmail.com
See page 885 for the editorial comment on this article (doi:10.1093/eurheartj/ehp106)
Aim: To date, most published echocardiographic methods have assessed left ventricular (LV) dyssynchrony (DYS) alone as a predictor for response to cardiac resynchronization therapy (CRT). We hypothesized that the response is instead dictated by multiple correctable factors.
Methods and results: A total of 161 patients (66 ± 10 years, EF 24 ± 6%, QRS > 120 ms) were investigated pre- and post-CRT (median of 6 months). Reduction in NYHA Class
1 or LV reverse remodelling (end-systolic volume reduction
10%) defined response. Four different pathological mechanisms were identified. Group1: LVDYS characterized by a pre-ejection septal flash (SF) (87 patients, 54%). Elimination of SF (77 of 87 patients) resulted in reverse remodelling in 100%. Group 2: short-AV delay (21 patients, 13%) resolution (19 of 21 patients) resulted in reverse remodelling in 16 of 19. Group 3: long-AV delay (16 patients, 10%) resolution (14 of 16 patients) resulted in NYHA Class reduction
1 in 11 with reverse remodelling in five patients. Group 4: exaggerated LV–RV interaction (15 patients, 9%) reduced post-CRT. All responded clinically with fall in pulmonary artery pressure (P = 0.003) but did not volume respond. Group 5: patients with none of the above correctable mechanisms (22 patients, 14%). None responded to CRT.
Conclusion: CRT response is dictated by correction of multiple independent mechanisms of which LVDYS is only one. Long-axis DYS measurements alone failed to detect 40% of responders.
Key Words: Echocardiography Heart failure Cardiac resynchronization therapy
This paper was guest edited by Prof. Martin J. Schalij, Department of Cardiology, Leiden University Medical Center, Leiden, The Netherlands
![]()
CiteULike
Connotea
Del.icio.us What's this?
Related articles in EHJ:
- Rocking will tell it
- Jens-Uwe Voigt
EHJ 2009 30: 885-886.[Extract] [Full Text]
This article has been cited by other articles:
![]() |
B. W.L. De Boeck, A. J. Teske, M. Meine, G. E. Leenders, M. J. Cramer, F. W. Prinzen, and P. A. Doevendans Septal rebound stretch reflects the functional substrate to cardiac resynchronization therapy and predicts volumetric and neurohormonal response Eur J Heart Fail, September 1, 2009; 11(9): 863 - 871. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-U. Voigt Rocking will tell it Eur. Heart J., April 2, 2009; 30(8): 885 - 886. [Full Text] [PDF] |
||||