European Heart Journal 1984 5(5):412-418;
Copyright © 1984 by the European Society of Cardiology.
This Article
Right arrow Full Text (PDF)
Right arrow E-letters: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when E-letters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by KUPARI, M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by KUPARI, M.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© 1984 The European Society of Cardiology

Reproducibility of M-mode echocardiographic assessment of left ventricular function. Significance of the temporal range of measurements

M. KUPARI

Cardiovascular Laboratory, Helsinki University Central Hospital, and Research Laboratories of the State Alcohol Monopoly (Alko) Helsinki, Finland

Received 19 May 1983; revised 19 October 1983; .

Address for correspondence Dr. Markku Kupari, MD, Cardiovascular Laboratory. Helsinki University Central Hospital, SF-00290 Helsinki 29, Finland.

Abstract

The aim of this study was to evaluate the reproducibility of M-mode echocardiography in sequential studies on left ventricular (LV) function. To that end, 3 to 8 serial recordings were performed on 4 groups of healthy subjects (total number, 49) within dissimilar time periods: (1) within 3 h, (2) within 24 h, (3) within 1 to 2 weeks and (4) within 14 to 18 months. The examination technique was carefully standardized and the recordings were analysed by digitization and computer assistance; the data were averaged from 5 beats on each recording. The random variation ofLV transverse dimensions and performance was roughly equal in the first 3 of these studies; the mean coefficients of variation ranged from 1.2 to 1.8% for end-diastolic dimension, from 1.7 to 2.3% for end systolic dimension, and from 2.5 to 3.6% for ejection fraction. In the long term study the respective mean coefficients were larger: 3.2%, 3.7% and 5.5%. The coefficients of variation for estimates of stroke volume, cardiac output, and peripheral vascular resistance ranged from 4.1 to 8.4% in studies completed within 1 day or inside Ior2 weeks but increased to 10-15.5% in the long-term study. The beat to beat variability of LV dimensions and performance, as assessed in a subgroup of 11 subjects, was roughly equal in quantity to the variability observed in the long-term study. It is concluded that provided the beat-to-beat variation is adequately taken into account the reproducibility of the basic M-mode echocardiographic LV measurements is excellent in short term studies in normal man and is satisfactory in studies repeated at intervals of several months or even longer. The use of this method to assess changes in cardiac volumes and haemodynamics should, however, be restricted to short-term studies.

Key Words: Reproducibility • echocardiography • left ventricular function


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
CirculationHome page
M. Kupari, A. Hautanen, L. Lankinen, P. Koskinen, J. Virolainen, H. Nikkila, and P. C. White
Associations Between Human Aldosterone Synthase (CYP11B2) Gene Polymorphisms and Left Ventricular Size, Mass, and Function
Circulation, February 17, 1998; 97(6): 569 - 575.
[Abstract] [Full Text] [PDF]



Disclaimer: Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.