European Heart Journal 1986 7(5):404-411;
Copyright © 1986 by the European Society of Cardiology.
This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrowRequest Permissions
Google Scholar
Right arrow Articles by JACK, C. M.
Right arrow Articles by ADGEY, A. A. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by JACK, C. M.
Right arrow Articles by ADGEY, A. A. J.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© 1986 The European Society of Cardiology

An external automatic device to detect ventricular fibrillation

C. M. JACK, E. K. HUNTER, T. H. PRINGLE, J. T. WILSON, J. ANDERSON and A. A. J. ADGEY

Regional Medical Cardiology Centre and University of Ulster, Royal Victoria Hospital Belfast BT12 6BA, U.K.

Received 7 October 1985; revised 30 December 1985; .

Abstract

An analysis of a micro-processor-based system to detect cardiac arrest rhythms was made in a series of 84 cardiac arrests in 78 patients. The ECG was sensed using 2 external ECGjdefibrillator pads applied to the chest wall. In 5 arrests, the initial rhythm was continuous electrical pacing in addition to the cardiac arrest and these were not included in the analysis. Of the remaining 79 arrests, there were 15 in which the initial rhythm was ventricular fibrillation and 14 were correctly detected by the system (sensitivity 93%). In the remaining 64 arrests the initial rhythm was not ventricular fibrillation and 62 were correctly identified (specificity 97%). A high specificity is required in any device employing automatic detection of ventricular fibrillation. Continuous electrocardiographic recordings of the cardiac arrests were analysed every 8 to 18 s: of 223 sections showing ventricular fibrillation, 165 were correctly determined by the system, i.e. sensitivity 74%, where the rhythm was not ventricular fibrillation, of the 5002 sections 4953 were correctly detected by the system, i.e. specificity 99%. The percentage accuracy of detection of ventricular fibrillation varied from 25–100% (mean 81%) and for detection of non-ventricular fibrillation from 57–100% (mean 99%) in every 8 to 18 s section of ECG tracing per cardiac arrest episode. Thus the automatic detection of cardiac arrest rhythms is feasible with a high degree of accuracy.

Key Words: Ventricular fibrillation • microprocessor • ECG/defibrillator pads


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?


This article has been cited by other articles:


Home page
CirculationHome page
F. Cecchin, D. B. Jorgenson, C. I. Berul, J. C. Perry, A. A. Zimmerman, B. W. Duncan, F. M. Lupinetti, D. Snyder, T. D. Lyster, G. L. Rosenthal, et al.
Is Arrhythmia Detection by Automatic External Defibrillator Accurate for Children? : Sensitivity and Specificity of an Automatic External Defibrillator Algorithm in 696 Pediatric Arrhythmias
Circulation, May 22, 2001; 103(20): 2483 - 2488.
[Abstract] [Full Text] [PDF]



Disclaimer:
Please note that abstracts for content published before 1996 were created through digital scanning and may therefore not exactly replicate the text of the original print issues. All efforts have been made to ensure accuracy, but the Publisher will not be held responsible for any remaining inaccuracies. If you require any further clarification, please contact our Customer Services Department.