European Heart Journal Advance Access originally published online on October 5, 2005
European Heart Journal 2006 27(2):157-165; doi:10.1093/eurheartj/ehi585
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Collateral-flow measurements in humans by myocardial contrast echocardiography: validation of coronary pressure-derived collateral-flow assessment
Department of Cardiology, University Hospital Bern, CH-3010 Bern, Switzerland
Received 13 June 2005; revised 28 July 2005; accepted 22 September 2005; online publish-ahead-of-print 5 October 2005.
* Corresponding author. Tel: +41 31 6323693; fax: +41 31 6324299. E-mail address: christian.seiler.cardio{at}insel.ch
See page 123 for the editorial comment on this article (doi:10.1093/eurheartj/ehi640)
| Abstract |
|---|
|
|
|---|
Aims Myocardial blood flow (MBF) is the gold standard to assess myocardial blood supply and, as recently shown, can be obtained by myocardial contrast echocardiography (MCE). The aims of this human study are (i) to test whether measurements of collateral-derived MBF by MCE are feasible during elective angioplasty and (ii) to validate the concept of pressure-derived collateral-flow assessment.
Methods and results Thirty patients with stable coronary artery disease underwent MCE of the collateral-receiving territory during and after angioplasty of 37 stenoses. MCE perfusion analysis was successful in 32 cases. MBF during and after angioplasty varied between 0.0600.876 mL min1 g1 (0.304±0.196 mL min1 g1) and 0.6761.773 mL min1 g1 (1.207±0.327 mL min1 g1), respectively. Collateral-perfusion index (CPI) is defined as the rate of MBF during and after angioplasty varied between 0.05 and 0.67 (0.26±0.15). During angioplasty, simultaneous measurements of mean aortic pressure, coronary wedge pressure, and central venous pressure determined the pressure-derived collateral-flow index (CFIp), which varied between 0.04 and 0.61 (0.23±0.14). Linear-regression analysis demonstrated an excellent agreement between CFIp and CPI (y=0.88x+0.01; r2=0.92; P<0.0001).
Conclusion Collateral-derived MBF measurements by MCE during angioplasty are feasible and proved that the pressure-derived CFI exactly reflects collateral relative to normal myocardial perfusion in humans.
Key Words: Coronary circulation Collateral circulation Collateral flow index Myocardial blood flow Myocardial perfusion Myocardial contrast echocardiography
| Introduction |
|---|
|
|
|---|
The coronary collateral circulation constitutes an alternative blood supply to myocardium jeopardized by ischaemia, and its protective role has been demonstrated by numerous investigations.1 Of considerable clinical relevance, coronary collateral arteries have become a target for novel treatment strategies by therapeutic arteriogenesis in patients with severe coronary artery disease (CAD).24 In order to demonstrate the efficacy of pro-arteriogenic substances, it is indispensable to quantitatively assess collateral flow, which in turn entails selective delineation of collateral pathways. The first technique fulfilling these requirements used coronary pressure measurements distal to a stenosis during angioplasty, i.e. coronary wedge pressure, which was related to angiographically visible collaterals.5 The current reference method measures collateral flow as a fraction of normal flow using a collateral-flow index (CFI), which is derived from simultaneous measurements of mean aortic pressure, coronary wedge pressure, and central venous pressure (CVP) during angioplasty. This CFI is theoretically well defined6 and has been confirmed by several invasive studies.79 Moreover, Matsuo et al.10 validated the index against perfusion defects, using 99 mTc-sestamibi during angioplasty. However, the direct verification of the CFI vs. the gold standard of myocardial blood supply assessment, that is, myocardial blood flow (MBF) (mL min1 g1), is still lacking. MBF, defined as blood flow (mL min1) into a region relative to its mass (g), can be obtained by positron emission tomography and lately by myocardial contrast echocardiography (MCE).11
The objectives of this human study are (i) to demonstrate that collateral-derived MBF can be obtained by MCE during elective angioplasty and (ii) to validate the concept of pressure-derived collateral-flow assessment by simultaneous measurements of coronary pressure and MBF.
| Methods |
|---|
|
|
|---|
Angioplasty model
Figure 1 depicts the flow model consisting of a myocardial area that is supplied by a stenotic coronary artery and an alternative pathway representing coronary collaterals. The biophysical properties of the myocardium are characterized by its flow resistance (Rmyo, mmHg min mL1) and mass (mmyo, g). The collateral-perfusion index (CPI, 1) is defined as collateral-derived MBF (MBFc, mL min1 g1) during balloon occlusion (left-hand panel) relative to normal MBF (MBFn, mL min1 g1) after removal of the stenosis (right-hand panel):
![]() | (1) |
|
Collateral-derived flow (Qc, ml min1) and normal coronary flow (Qn, ml min1) are controlled by their perfusion pressure, i.e. coronary wedge pressure (Poccl, mmHg) during angioplasty and aortic pressure (Pao, mmHg) after successful angioplasty, and the CVP (mmHg), which acts as backpressure:
![]() | (2) |
![]() | (3) |
The pressure-derived CFI (CFIp, 1) equals:
![]() | (4) |
Study population and protocol
For this pilot study, we planned to investigate 30 patients with stable CAD eligible for percutaneous coronary intervention (PCI) of
1 stenotic lesion. Forty-four consecutive patients (
18 years) referred for elective coronary angiography were initially evaluated to participate in the study. Six patients refused consent and eight patients were excluded after the angiographic assessment. The patients were not screened for echocardiographic image quality and maintained their routine medication.
All measurements were performed in the catheterization laboratory during one session with the patient in supine position lying on his back. After diagnostic coronary angiography, constant venous ultrasound contrast agent (UCA) infusion was started and the myocardial territory of the coronary artery undergoing angioplasty, i.e. the collateral-receiving artery, was imaged by transthoracic MCE. PCI was performed after stable myocardial enhancement was reached. During balloon occlusion, myocardial perfusion sequences were captured while continuously recording coronary wedge pressure, aortic pressure, and CVP. After successful completion of the intervention, that is, restoration of normal blood flow at rest, perfusion imaging of the same territory was repeated following cessation of reactive hyperaemia. MCE und pressure studies were performed and analysed by two independent examiners blinded to the other methods' results. The study protocol was approved by the Ethics Committee of the University of Bern, and all participants gave written informed consent to participate in the study.
Cardiac catheterization and pressure measurements
After two puffs of isosorbide dinitrate, diagnostic coronary angiography was performed via the right femoral artery approach, using five French Judkins catheters. Indication for ad hoc PCI is based on the visual estimate of the stenosis per cent diameter reduction. For PCI and coronary pressure measurements, a six French-guiding catheter and a 0.014 in. guidewire equipped with a pressure sensor at the tip (PressureWire, RADI Medical, Uppsala, Sweden) was used. Simultaneous measurements of mean aortic pressure through the angioplasty-guiding catheter, coronary wedge pressure distal to the stenosis undergoing PCI, and CVP via a femoral venous access were performed during balloon occlusion. Stenosis severity was quantitatively assessed offline as per cent diameter reduction of the vessel (Philips DA, Best, The Netherlands).
Myocardial contrast echocardiography
Ultrasound contrast agent
UCA was administered via the right cubital vein, using parallel infusion of FS069 (OPTISON, Amersham Health SA, Oslo, Norway) 3 mL at a rate of 1030 mL h1 (Perfusor fm, B. Braun Melsungen AG, Melsungen, Germany) and physiological saline at a rate of 400 mL h1 (VOLUMED®
VP2001, arcomed ag, Regensdorf, Switzerland). The UCA infusion pump was manually agitated to guarantee UCA homogenization.
Data acquisition
Transthoracic imaging was performed with an Acuson Sequoia C256 ultrasound scanner (Siemens Medical Solutions, Mountain View, CA, USA) equipped with a 3V2c transducer and Coherent Contrast Imaging®. Settings were as follows: MI for microsphere detection 0.08, MI for microsphere destruction 1.3, dynamic range 60 dB, and linear post-processing. The image plane was adjusted to the myocardial territory of the coronary artery to be dilated, i.e. the collateral-receiving territory. Gain was adjusted for optimal visualization of the myocardium and held constant throughout the image acquisition. Destruction-refill sequences were generated using the manual bubble destruction (MBD) feature of the scanner and recorded digitally for offline image analysis. The clip length was set to 200 frames with a triggered interval of 75 ms, resulting an acquisition time of 15 s. Prior to MBD, two cardiac cycles were captured for the calculation of the relative blood volume.
Data analysis
Image visualization and quantification was done with DataPro 2.11 (Noesis S.A., Courtaboeuf, France). Logarithmic signal compression was removed, and linearized signal intensity data were expressed in arbitrary units. Quantitative perfusion analysis was performed on end-systolic frames selected from the perfusion sequence. Appropriate regions of interest were placed and tracked manually within the myocardium and an adjacent region within the left ventricular cavity.11 Myocardial intensity data were corrected for non-contrast signals arising from the tissue by subtracting the signal intensity of the first frame after MBD. Myocardial plateau intensity A was calculated by averaging myocardial signal intensity data from frames before MBD (Figure 2). Parameter ß (min1) was derived from the fitting of myocardial intensity data after MBD to the refill equation [Eq. (5)].
|
![]() | (5) |
Averaging left ventricular signal intensity data of all but the frames during and the first one after MBD yielded ALV and the relative blood volume (rBV, mL mL1):
![]() | (6) |
MBF was calculated according to Eq. (7) with a tissue density
T of 1.05 g mL1:
![]() | (7) |
Statistical methods
Demographic and clinical data were expressed as mean±SD. Paired, two-sided t-tests were used for the comparison of pressure and perfusion data. Correlations were performed using linear-regression analysis, and the accuracy of the prediction was measured by the standard error of the estimate. Statistical significance was defined at a probability value P<0.05. The Bonferroni correction was used for multiple comparisons between CPI and the pressure indices Poccl/Pao and CFIp. Agreement between two measurement techniques was established using BlandAltman analysis by plotting the differences between the two techniques against the mean values of the two techniques; values of mean±2SD of measurement differences were added to the plots. The range within which the measurement differences were required to lie was not pre-defined.
| Results |
|---|
|
|
|---|
The past medical history of the 30 patients revealed stable angina in 25 cases. Table 1 summarizes patient's characteristics and angiographic data. Left ventricular ejection fraction and end-diastolic pressure (LVEDP) ranged between 30 and 70% and 3 and 1 mmHg, respectively. PCI was performed in 37 stenotic coronary arteries. The per cent diameter reduction ranged between 30 and 100%, including seven non-significant stenoses. Angioplasty of non-significant stenoses was performed because of visual overestimation in six cases (4149% diameter reduction by quantitative analysis) and because of an ulcerated plaque in one case (30% diameter reduction). During angioplasty of 37 stenoses, angina was present in 19 and ST-segment changes on the surface ECG in 20 cases. Invasive pressure measurements were obtained from all lesions (Table 2). MCE perfusion analysis during and after angioplasty was completed in 32 territories (Table 2) and failed in one lateral and four inferior territories. The causes for unsuccessful data acquisition of the four inferior territories were premature study termination due to an intervention-related serious adverse event in one patient, technical problems in two cases (scanner dysfunction and lack of synchronicity between sonographer and interventionalist), and poor image quality in one patient with severe chronic obstructive pulmonary disease.
|
|
Figure 2 summarizes perfusion sequences and analyses during and after PCI of a left anterior descending coronary artery stenosis with a 52% diameter reduction and a pressure-derived CFI of 0.17. Visual comparison of the perfusion sequences suggests modest collateralization indicated by decelerated UCA refill during balloon inflation. MBF during and after successful angioplasty was 0.161 mL min1g1 and 0.823 mL min1 g1, respectively, resulting in a CPI of 0.20. Although relative myocardial blood volumes during and after PCI were similar (0.088 mL mL1 vs. 0.080 mL mL1), the corresponding blood volume exchange frequency ß increased from 1.920 to 10.740 min1.
The graphical and statistical analyses of absolute (Poccl, Poccl-CVP) and relative (Poccl/Pao, CFIp) pressure data vs. absolute (MBFc) and relative (CPI) perfusion data are shown in Figures 3 and 4, respectively. Coronary wedge pressure, effective coronary perfusion pressure (Poccl-CVP), coronary wedge pressure relative to aortic pressure (Poccl/Pao) as well as CFI, CFIp, correlated linearly with absolute collateral-derived myocardial perfusion, and collateral-derived relative to normal myocardial perfusion. In accordance with the theoretical concept, the best agreement was found between pressure-derived CFI and CPI.
|
|
The measurement differences between non-dimensional collateral-pressure and -perfusion indexes, i.e. Poccl/Pao, CFIp, and CPI, are depicted in Figure 5. Averaged over the entire range, the CPI was overestimated by Poccl/Pao (0.030±0.053, P=0.003) and underestimated by CFIp (0.023±0.045, P=0.007). BlandAltman plots suggest proportional errors between Poccl/Pao and CPI (Figure 5A) as well as, but less apparently, between CFIp and CPI (Figure 5B).
|
In our study population, the measurement difference (CPICFIp) decreased linearly with increasing LVEDP (Figure 6); CFIp underestimated CPI for filling pressures up to 16 mmHg and overestimated CPI for higher filling pressures.
|
| Discussion |
|---|
|
|
|---|
To date, coronary collateral flow has been assessed by several qualitative and quantitative methods and has been compared with clinical markers of collateralization and prognostic outcome in patients with CAD.1 This study, for the first time, demonstrates that collateral-derived MBF can be measured using MCE during angioplasty and, in comparison with simultaneous coronary pressure measurements, confirms the theoretical concept of pressure-derived collateral-flow assessment in humans. In this context and for future practical application, the question arises what would be the ideal parameter to characterize the coronary collateral circulation and whether it should be assessed in absolute terms or relative to the flow in the normally patent coronary artery.
MCE and the assessment of the collateral circulation
MCE has been proposed for the assessment of collateral-derived myocardial perfusion by several authors.1216 Two human studies fulfilling the mandatory requirements, i.e. documented total coronary occlusion to avoid concomitant contrast flow via the native vessel, compared MCE and invasive collateral assessment. In patients with recent acute myocardial infarction,13 angiographically visible collaterals correlated poorly with the size of the collateralized area as well as normalized UCA transit rates. In patients with stable CAD undergoing PCI,15 pressure-derived CFI was modestly associated with peak signal intensity of UCA transit curves but not with UCA transit rates. The conflicting findings regarding UCA transit rates may be explained by the normalization performed by Sabia et al.13 However, the reliability of UCA transit curve parameters is limited because of their dependency on contrast agent concentration and mode of bolus application. Furthermore, the studies differed by the setting (acute myocardial infarction vs. stable CAD) and the reference method (visual13 vs. pressure-derived15 collateral assessment).
Refill curve parameter ß has been shown to represent MBF in anaesthetized dogs.17 However, the comparison with PET revealed only modest agreement between segmental MBF and ß-data in conscious humans.11 Likewise, the present study demonstrated weak correlation between Poccl and parameter ß during angioplasty (r2=0.16, P=0.02, data not shown) as well as between CFIp and the ratio of parameter ß during and after angioplasty (r2=0.18, P=0.01, data not shown).
The methods discussed earlier share a common characteristic, which limits their applicability. The collateral-derived blood supply was described either by the collateral-dependent blood content (peak signal video intensity and size of the collateralized area) or by the exchange rate of this blood content (UCA transit rate, parameter ß). Our method overcomes this drawback as the collateral-dependent blood volume and its exchange frequency are combined to MBF given in mL min1 g1 [Eq. (7)]. With respect to the quantitative evaluation of the collateralization by MCE, we therefore suggest to assess collateral derived relative to normal myocardial perfusion on the basis of MBF measurements during and after angioplasty. MBF measurements by MCE have recently been introduced and validated against PET and intracoronary Doppler in healthy volunteers and patients with CAD.11 Moreover, as single photon emission computed tomography10 cannot provide MBF and PET is not applicable because of long acquisition times requiring coronary occlusion >5 min, MBF measurements by MCE have to be recognized as the gold standard to quantify the collateral-derived myocardial blood supply.
Absolute parameters describing the collateral circulation
Presence or absence of angina pectoris and ECG ST-segment changes during balloon occlusion are related to the extent of collateralization.7,18 However, the diagnostic value of angina pectoris is rather qualitative and the ECG's sensitivity to detect alterations in collateralization has not yet been defined.
At present, quantitative approaches to the human coronary collateralization rely on parameters such as X-ray contrast media clearance, radionuclide activity, and most notably, coronary pressure and flow velocity.1
It has been shown that absolute distal occlusive collateral-flow velocity as determined by intracoronary Doppler moderately correlated with coronary wedge pressure19 and coronary wedge pressure relative to aortic pressure.20 Quantitative assessment of 99 mTc-sestamibi perfusion defects during angioplasty in patients with CAD further supported the association between coronary wedge pressure and collateral-derived myocardial blood supply.10 The results of the present study finally establish the agreement between coronary wedge pressure and collateral-derived myocardial blood supply (Figure 3A).
In summary, these findings indicate that absolute coronary wedge pressure is a valuable indicator of collateral-derived blood supply, whereas absolute distal occlusive coronary flow velocity is less reliable. This result may be explained by inter-individual biological variability of local coronary artery cross-sectional areas21 and flow velocities, which finally constitute coronary blood flow,22 as well as by better reliability and superior robustness of pressure when compared with Doppler flow measurements.
Relative parameters describing the collateral circulation
Human studies on myocardial perfusion reported considerable heterogeneity of normal MBF data.11,23 Thus, indices describing collateral relative to normal myocardial blood supply are expected to be more suitable for determining the relevance of collateral vessels. Consistently, the accuracy of intracoronary Doppler measurements has been improved by normalizing coronary flow velocity taken at the same location after the removal of the stenosis and following cessation of reactive hyperaemia.7,19
By normalizing coronary wedge pressure with aortic pressure, Matsuo et al.10 demonstrated improved agreement between pressure-derived collateral-flow data and radionuclide uptake defects during angioplasty. In our patients, relative and absolute pressure data correlated better with relative than with absolute collateral perfusion (Figures 3 and 4). In other words, the scatter between pressure- and perfusion-derived collateral data mainly emerges from inter-individual variations of normal perfusion, whereas aortic and CVP vary within smaller limits. Thus, normalization with MBF may partly explain the improved consistency of our data when compared with the results from Matsuo et al.10 Besides, our method features two methodological advantages. First, the biophysical concept of pressure-derived collateral-flow measurements is based on considerations of myocardial perfusion. Secondly, perfusion imaging by MCE was completed during vessel occlusion and was not impaired by delayed scanning procedures due to the required transfer from cardiac catheterization laboratory to nuclear imaging facility.
Likely unexpected for some researchers in the field,20,24 this cross-sectional study supports the consideration of CVP in order to increase the accuracy of pressure-derived collateral-flow data. Backpressure compensation is even more important in the case of follow-up studies, e.g. trials on therapeutic arteriogenesis, where it is desired to uncouple collateral-flow data from changing haemodynamic parameters. We therefore recommend in agreement with the theory6 to include CVP as well as aortic pressure for the quantitative assessment of the collateralization by invasive pressure measurements.
For the pressureflow relationships [Eqs. (2) and (3)] of the angioplasty model, we assumed that coronary blood vessels are rigid tubes, i.e. wall stress has no influence on the coronary circulation. However, coronary vessels are elastic and external pressures affect their haemodynamic properties. LVEDP is recognized as a useful substitute for wall stress, and it has been shown that collateral flow was overestimated by pressure when compared with Doppler flow measurements in the case of high LVEDP.25 Correspondingly, we found that CFIp overestimated CPI for LVEDP >16 mmHg. This phenomenon is called coronary waterfall: the volume flowrate in a waterfall depends on the conditions at the top of the fall and is independent of the height of the drop.26
Non-invasive assessment of the collateral circulation
MCE has been proposed to become the optimal tool in the area of therapeutic angiogenesis.14 Using the quantitative approach as presented in this study, MCE can now be recognized as the reference method for the non-invasive quantification of coronary flow. However, we have to stress that invasive access is still required either to selectively delineate collateral pathways by brief artificial vessel occlusion or to confirm total coronary occlusion.1 Future developments such as targeted microspheres27 or highly selective UCA destruction will show whether a pure non-invasive approach to the collateral function will be feasible by MCE. For the present, MCE in conjunction with our quantification algorithm is ready for clinical studies on coronary collaterals in patients with documented total coronary occlusion.
Limitations
We investigated 37 coronary arteries with their territories in 30 patients and independence of the data, a prerequisite for the statistical analysis, may be challenged. From the technical viewpoint, the use of coronary arteries as the basic unit for the statistical analyses is reasonable. Furthermore, the analysis considering only the first vessel/territory per patient (data not shown) revealed only marginal changes in haemodynamic statistics and did not change the significance of any result.
As a tomographic technique, the following deficiencies that may influence the results are inherent in MCE: although we took every precaution to locate the same myocardial territory during and after angioplasty, we cannot quantify the measurement error related to the investigation of unequal tomographic planes. Furthermore, MCE cannot assess collateral-related processes beyond the UCA detection volume of the transducer.
In some patients, collateral-derived MBF was below the range covered by the human validation study.11 However, there are no theoretical restrictions regarding the analysis of slow refill curves.
Finally, we cannot comprehensively clarify the slight underestimation of CPI by CFIp. It may be caused by changing haemodynamic conditions between the simultaneous assessment of MBFc and CFIp during angioplasty and the perfusion follow-up after angioplasty or by MBFn reduction due to peri-procedural embolization into the microcirculation.
| Conclusion |
|---|
|
|
|---|
During angioplasty, measurements of collateral-derived MBF by contrast echocardiography are feasible and proved the concept of coronary pressure-derived collateral-flow assessment in humans. MBF measurements by contrast echocardiography represent an important step towards the non-invasive quantification of collateral function and are ready to be used for clinical studies in models with documented total vessel occlusion.
| Acknowledgement |
|---|
|
|
|---|
This work was supported by grants from the Swiss National Science Foundation (no. 3200B0-100065/1, C.S.) and the Swiss Heart Foundation.
Conflict of interest: the authors have no conflict of interest to disclose.
| References |
|---|
|
|
|---|
- Seiler C. The human coronary collateral circulation. Heart 2003;89:13521357.
[Free Full Text] - Schaper W, Ito WD. Molecular mechanisms of coronary collateral vessel growth. Circ Res 1996;79:911919.
[Free Full Text] - Carmeliet P. Mechanisms of angiogenesis and arteriogenesis. Nature Med 2000;6:389395.[CrossRef][Web of Science][Medline]
- Seiler C, Pohl T, Wustmann K, Hutter D, Nicolet PA, Windecker S, Eberli FR, Meier B. Promotion of collateral growth by granulocytemacrophage colony-stimulating factor in patients with coronary artery disease. A randomized, double-blind, placebo-controlled study. Circulation 2001;104:20122017.
[Abstract/Free Full Text] - Meier B, Luethy P, Steffenino GD, Rutishauser W. Coronary wedge pressure in relation to spontaneously visible and recruitable collaterals. Circulation 1987;75:906913.
[Abstract/Free Full Text] - Pijls NH, van Son JA, Kirkeeide RL, de Bruyne B, Gould KL. Experimental basis of determining maximum coronary, myocardial, and collateral blood flow by pressure measurements for assessing functional stenosis severity before and after percutaneous transluminal coronary angioplasty. Circulation 1993;87:13541367.
[Abstract/Free Full Text] - Seiler C, Fleisch M, Garachemani A, Meier B. Coronary collateral quantitation in patients with coronary artery disease using intravascular flow velocity or pressure measurements. J Am Coll Cardiol 1998;32:12721279.
[Abstract/Free Full Text] - Seiler C, Billinger M, Fleisch M, Meier B. Washout collaterometry: a new method of assessing collaterals using angiographic contrast clearance during coronary occlusion. Heart 2000;86:540546.
- Werner GS, Ferrari M, Heinke S, Kuethe F, Surber R, Richartz BM, Figulla HR. Angiographic assessment of collateral connections in comparison with invasively determined collateral function in chronic coronary occlusions. Circulation 2003;107:19721977.
[Abstract/Free Full Text] - Matsuo H, Watanabe S, Kadosaki T, Yamaki T, Tanaka S, Miyata S, Segawa T, Matsuno Y, Tomita M, Fujiwara H. Validation of collateral fractional flow reserve by myocardial perfusion imaging. Circulation 2002;105:10601065.
[Abstract/Free Full Text] - Vogel R, Indermühle A, Reinhardt J, Meier P, Siegrist PT, Namdar M, Kaufmann PA, Seiler C. The quantification of absolute myocardial perfusion in humans using contrast echocardiography: algorithm and validation. J Am Coll Cardiol 2005;45:754762.
[Abstract/Free Full Text] - Grill HP, Brinker JA, Taube JC, Walford GD, Midei MG, Flaherty JT, Weiss JL. Contrast echocardiographic mapping of collateralized myocardium in humans before and after angioplasty. J Am Coll Cardiol 1990;16:15941600.[Abstract]
- Sabia PJ, Powers ER, Jayaweera AR, Ragosta M, Kaul S. Functional significance of collateral blood flow in patients with recent acute myocardial infarction. A study using myocardial contrast echocardiography. Circulation 1992;85:20802089.
[Abstract/Free Full Text] - Mills JD, Fischer D, Villanueva FS. Coronary collateral development during chronic ischaemia: serial assessment using harmonic myocardial contrast echocardiography. J Am Coll Cardiol 2000;36:618624.
[Abstract/Free Full Text] - de Marchi SF, Schwerzmann M, Fleisch M, Billinger M, Meier B, Seiler C. Quantitative contrast echocardiographic assessment of collateral derived myocardial perfusion during elective coronary angioplasty. Heart 2001;86:324329.
[Abstract/Free Full Text] - Coggins MP, Sklenar J, Le DE, Wei K, Lindner JR, Kaul S. Noninvasive prediction of ultimate infarct size at the time of acute coronary occlusion based on the extent and the magnitude of collateral derived myocardial blood flow. Circulation 2001;104:24712477.
[Abstract/Free Full Text] - Wei K, Jayaweera AR, Firoozan S, Linka A, Skyba DM, Kaul S. Quantification of myocardial blood flow with ultrasound-induced destruction of microbubbles administered as a continuous venous infusion. Circulation 1998;97:473483.
[Abstract/Free Full Text] - Pijls NH, Bech GJ, El Gamal MI, Bonnier HJ, de Bruyne B, van Gelder B, Michels HR, Koolen JJ. Quantification of recruitable coronary collateral blood flow in conscious humans and its potential to predict future ischemic events. J Am Coll Cardiol 1995;25:15221528.[Abstract]
- Mohri M, Egashira K, Kuga T, Shimokawa H, Takeshita A. Correlations between recruitable coronary collateral flow velocities, distal occlusion pressure, and electrocardiographic changes in patients undergoing angioplasty. Jpn Circ J 1997;61:971978.[Medline]
- van Liebergen RA, Piek JJ, Koch KT, de Winter RJ, Schotborgh CE, Lie KI. Quantification of collateral flow in humans: a comparison of angiographic, electrocardiographic and hemodynamic variables. J Am Coll Cardiol 1999;33:670677.
[Abstract/Free Full Text] - MacAlpin RN, Abbasi AS, Grollman JH, Eber L. Human coronary artery size during life. A cinearteriographic study. Radiology 1978;108:567576.
- Anderson HV, Stokes MJ, Leon M, Abu-Halawa SA, Stuart Y, Kirkeeide RL. Coronary artery flow velocity is related to lumen area and regional left ventricular mass. Circulation 2000;102:4854.
[Abstract/Free Full Text] - Chareonthaitawee P, Kaufmann PA, Rimoldi O, Camici PG. Heterogeneity of resting and hyperemic myocardial blood flow in healthy humans. Cardiovasc Res 2001;50:151161.
[Abstract/Free Full Text] - Werner GS, Ferrari M, Heinke S, Kuethe F, Surber R, Richartz BM, Figulla HR. Angiographic assessment of collateral connections in comparison with invasively determined collateral function in chronic coronary occlusions. Circulation 2003;107:19721977.
- de Marchi SF, Oswald Ph, Windecker S, Meier B, Seiler C. Reciprocal relationship between left ventricular filling pressure and the recruitable human coronary collateral circulation. Eur Heart J 2005;26:558566.
[Abstract/Free Full Text] - Fung YC. Biomechanics: Circulation. 2nd ed. New York: Springer; 1997. p507509.
- Leong-Poi H, Christiansen J, Klibanov AL, Kaul S, Lindner JR. Noninvasive assessment of angiogenesis by ultrasound and microbubbles targeted to alpha(v)-integrins. Circulation 2003;107:455460.
[Abstract/Free Full Text]
Related articles in EHJ:
- Assessment of the collateral circulation of the heart
- Nico H.J. Pijls
EHJ 2006 27: 123-124.[Extract] [FREE Full Text]
This article has been cited by other articles:
![]() |
P Meier, J Antonov, R Zbinden, A Kuhn, S Zbinden, S Gloekler, M Delorenzi, R Jaggi, and C Seiler Non-invasive gene-expression-based detection of well-developed collateral function in individuals with and without coronary artery disease Heart, June 1, 2009; 95(11): 900 - 908. [Abstract] [Full Text] [PDF] |
||||
![]() |
R Vogel, A Indermuhle, P Meier, and C Seiler Quantitative stress echocardiography in coronary artery disease using contrast-based myocardial blood flow measurements: prospective comparison with coronary angiography Heart, March 1, 2009; 95(5): 377 - 384. [Abstract] [Full Text] [PDF] |
||||
![]() |
S H Schirmer, F C van Nooijen, J J Piek, and N van Royen Stimulation of collateral artery growth: travelling further down the road to clinical application Heart, February 1, 2009; 95(3): 191 - 197. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. H. Schirmer, J. O. Fledderus, P. T.G. Bot, P. D. Moerland, I. E. Hoefer, J. Baan Jr, J. P.S. Henriques, R. J. van der Schaaf, M. M. Vis, A. J.G. Horrevoets, et al. Interferon-{beta} Signaling Is Enhanced in Patients With Insufficient Coronary Collateral Artery Development and Inhibits Arteriogenesis in Mice Circ. Res., May 23, 2008; 102(10): 1286 - 1294. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Meier, S. Gloekler, R. Zbinden, S. Beckh, S. F. de Marchi, S. Zbinden, K. Wustmann, M. Billinger, R. Vogel, S. Cook, et al. Beneficial Effect of Recruitable Collaterals: A 10-Year Follow-Up Study in Patients With Stable Coronary Artery Disease Undergoing Quantitative Collateral Measurements Circulation, August 28, 2007; 116(9): 975 - 983. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. di Mario and K. Dimopoulos The Effect of Drug-Eluting Stents on Collateral Coronary Flow J. Am. Coll. Cardiol., August 7, 2007; 50(6): 560 - 560. [Full Text] [PDF] |
||||
![]() |
L. O. Jensen, P. Thayssen, J. F. Lassen, H. S. Hansen, H. Kelbaek, A. Junker, K. E. Pedersen, K. N. Hansen, L. R. Krusell, H. E. Botker, et al. Recruitable collateral blood flow index predicts coronary instent restenosis after percutaneous coronary intervention Eur. Heart J., August 1, 2007; 28(15): 1820 - 1826. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Chalothorn, J. A. Clayton, H. Zhang, D. Pomp, and J. E. Faber Collateral density, remodeling, and VEGF-A expression differ widely between mouse strains Physiol Genomics, July 18, 2007; 30(2): 179 - 191. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Berry, K. P. Balachandran, P. L. L'Allier, J. Lesperance, R. Bonan, and K. G. Oldroyd Importance of collateral circulation in coronary heart disease Eur. Heart J., February 1, 2007; 28(3): 278 - 291. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Meier, R. Zbinden, M. Togni, P. Wenaweser, S. Windecker, B. Meier, and C. Seiler Coronary Collateral Function Long After Drug-Eluting Stent Implantation J. Am. Coll. Cardiol., January 2, 2007; 49(1): 15 - 20. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. J. Bax, B. De Bruyne, A. K. Gitt, S. Kristensen, C. Linde, D. Poldermans, F. J. Pinto, P. Ponikowski, B. D. Prendergast, E. Abagiti-Rosei, et al. Highlights of the 2006 Scientific Sessions of the European Society of Cardiology: Barcelona, Spain, September 2-5, 2006 J. Am. Coll. Cardiol., December 19, 2006; 48(12): 2564 - 2574. [Full Text] [PDF] |
||||
![]() |
J. Spaan and M. Siebes Is myocardial contrast echocardiography ready to assume 'gold standard' status for quantification of collateral flow in humans? Eur. Heart J., July 1, 2006; 27(13): 1627 - 1627. [Full Text] [PDF] |
||||
![]() |
R. Vogel, R. Zbinden, A. Indermuhle, S. Windecker, B. Meier, and C. Seiler Is myocardial contrast echocardiography ready to assume 'gold standard' status for quantification of collateral flow in humans?: reply Eur. Heart J., July 1, 2006; 27(13): 1627 - 1628. [Full Text] [PDF] |
||||
![]() |
N. H.J. Pijls Assessment of the collateral circulation of the heart Eur. Heart J., January 2, 2006; 27(2): 123 - 124. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||






) and myocardial () signal intensities of the mid-segment are plotted against time. Upper and lower dotted lines indicate mean left ventricular signal intensity ALV and myocardial plateau signal intensity A, respectively.