European Heart Journal Advance Access originally published online on October 11, 2006
European Heart Journal 2006 27(21):2567-2572; doi:10.1093/eurheartj/ehl303
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Assessment of coronary artery stent restenosis by 64-slice multi-detector computed tomography
1 Department of Internal Medicine 2 (Cardiology), University of Erlangen, Ulmenweg 18, 91054 Erlangen, Germany
2 Institute for Diagnostic Radiology, University of Erlangen, Erlangen, Germany
Received 2 May 2006; revised 7 September 2006; accepted 21 September 2006; online publish-ahead-of-print 11 October 2006.
* Corresponding author. Tel: +49 9131 8535000; fax: +49 9131 8535303. E-mail address: stephan.achenbach{at}med2.med.uni-erlangen.de
| Abstract |
|---|
|
|
|---|
Aims We investigated the feasibility of assessing coronary artery stent restenosis using a new generation 64-slice multi-detector computed tomography-scanner (MDCT) in comparison to conventional quantitative angiography.
Methods and results MDCT was performed in 64 consecutive patients (mean age 58±10 years) with previously implanted coronary artery stents (102 stented lesions: mean stent diameter 3.17±0.38 mm). Each stent was classified as evaluable or unevaluable, and in evaluable stents, the presence of in-stent restenosis (diameter reduction >50%) was determined visually. Results were verified against invasive, quantitative coronary angiography. Fifty-nine stented lesions (58%) were classified as evaluable in MDCT. The mean diameter of evaluable stents was 3.28±0.40 mm, whereas the mean diameter of non-evaluable stents was 3.03±0.31 mm (P=0.0002). Overall, six of 12 in-stent restenoses were correctly detected by MDCT [50% sensitivity (confidence interval 2277%)] and in 51 of 90 lesions, in-stent restenosis was correctly ruled out [57% specificity (4667%)]. In evaluable stents, six of seven in-stent restenoses were correctly detected, and the absence of in-stent stenosis was correctly identified in 51 of 52 cases [sensitivity 86% (4299%) and specificity 98% (88100%)].
Conclusion Stent type and diameter influence evaluability concerning in-stent restenosis by MDCT. The rate of assessable stents is low, but in evaluable stents, accuracy for detection of in-stent restenosis can be high.
Key Words: Computed tomography Stent Restenosis Angiography
| Introduction |
|---|
|
|
|---|
The clinical incidence of restenosis after coronary stent implantation is
2035% for bare metal stents and 510% for drug-eluting stents,15 but it can be higher in certain subsets of lesions such as long stenoses, bifurcation lesions, or lesions in small coronary arteries.68 Given the high number of patients who receive coronary artery stents, a non-invasive tool for the reliable detection of in-stent restenosis would be clinically useful. Standard non-invasive techniques, such as stress testing, lack sufficient sensitivity and specifity.9 Thus, there has been growing interest in the use of multi-detector computed tomography (MDCT) for the assessment of coronary artery stents. Sixteen-slice and 64-slice MDCT have been shown to permit the detection of coronary artery stenoses in the native coronary arteries with sensitivities and specificities up to 99%,1020 but did not show sufficient spatial and temporal resolution for the detection of in-stent restenosis. In five studies performed with 16- and 40-slice MDCT and including 42 to 232 stents, up to 77% of stents were considered unevaluable and even in the evaluable stents, sensitivities and specificities for the detection of in-stent restenosis were as low as 54%.2125 Two studies (29 and 75 patients) investigated the use of 16- and 64-slice CT for the assessment of stents in the left main coronary artery and found lower rates of unevaluable stents (7 and 5%).26,27 In both studies, all in-stent restenoses were detected by MDCT. The mean diameter of the stents was 3.9 mm in both studies, which indicated that the limited spatial resolution of 16-slice MDCT may be one of the reasons for the low number of evaluable stents in the other coronary arteries. Sixty-four-slice MDCT has higher spatial and temporal resolution than previous scanner generations, but its accuracy for detection of in-stent restenosis has not been reported. We therefore investigated 64 consecutive patients with 102 stented lesions to systematically evaluate the method's potential to assess coronary artery stents of various types and sizes concerning the presence of in-stent restenosis.
| Methods |
|---|
|
|
|---|
Sixty-four consecutive patients with previous stent implantation (102 stented lesions) who were scheduled for invasive coronary angiography were included in the study. If two stents were contiguously implanted in one lesion, they were considered as one single stent (in all lesions with two implanted stents, the stents were of equal type and diameter).
Forty-one patients were male and 23 were female, the mean age was 58±10 years. Table 1 lists details of the implanted stents. The mean time interval from stent implantation to inclusion in the study was 13.5±16.4 months (470 months).
|
Stent-in-stent implantation and renal failure, known allergy to contrast agent, possible pregnancy, and non-sinus rhythm were exclusion criteria for the study. All coronary artery stents in native or coronary bypass vessels were included, and stent type and nominal stent diameter were documented. All patients gave written informed consent to the study procedure and the study was approved by the institutional review board.
Patients with a heart rate greater than 60 b.p.m. received 100 mg of atenolol orally 45 min before MDCT. If the heart rate remained above 60 b.p.m. at the time of the MDCT scan, up to four doses of 5 mg metoprolol were given intravenously. Mean heart rate during the MDCT scan was 60±5 b.p.m. (range: 4964 b.p.m.). All patients received a total dose of 0.8 mg isosorbide dinitrate sublingually before the MDCT examination.
Images were acquired using a 64-slice MDCT scanner (Somatom Sensation 64, Siemens Medical Solutions, Forchheim, Germany), with 64x0.6 mm collimation, 330 ms rotation time, and a table feed of 3.8 mm per rotation. The contrast agent transit time was measured using a test bolus injection (10 mL contrast agent, followed by 50 mL saline solution, all at a flow rate of 5 mL/s) as reported previously.19 For visualization of the coronary arteries and stents, 5575 mL of contrast agent (370 mg iodine/ml) was injected intravenously at 5 mL/s, followed by 50 mL of saline solution. Tube voltage was 120 kV and effective tube current was 850 mAs. Tube current modulation was used in 34 patients with a heart rate of <60 b.p.m. to reduce radiation exposure during systole. In patients with a heart rate of greater 60 b.p.m., tube current modulation was deactivated to allow maximal flexibility in chosing image reconstruction intervals.
Reconstructions were obtained at 70% of the RR interval using a sharp reconstruction kernel (B46f), with a slice thickness of 0.6 mm and increment of 0.3 mm. If the data set showed motion artefact, additional reconstructions were performed in 5% decrements (65%, 60%, 55%, ...) until a data set free of motion artefact was obtained. This included data sets reconstructed in systole, if diastolic data sets showed motion artefact. If no data set was entirely free of motion artefact, the data set with best image quality in the region of the implanted stent was used for further analysis.
MDCT data sets were analysed on dedicated workstations (Leonardo®, Siemens Medical Solutions). Review of the original transaxial images, oblique multi-planar reconstructions (MPRs), and curved MPR were used to evaluate the implanted coronary stents concerning the presence of in-stent restenosis (Figure 1). Two readers experienced in cardiac MDCT and blinded to the patients' clinical history as well as the stent type and dimensions jointly analysed all implanted coronary artery stents in consensus. In a first step, each stent was classified as evaluable or unevaluable. A stent was classified as unevaluable if blurring of the stent contours or image noise prevented clear separation of the metal stent structure and the in-stent lumen. In a second step, evaluable stents were visually classified as to the presence or absence of in-stent restenosis (diameter reduction
50%).
|
Invasive coronary angiography was performed 120 days after MDCT scanning. After intracoronary injection of 0.2 mg of isosorbide dinitrate, standardized projections showing the stented segment in at least two projections were obtained. Invasive angiograms were evaluated off-line using quantitative coronary angiography (QCA) software (QuantCor®, Siemens Medical Solutions) by an experienced observer unaware of MDCT results. A significant in-stent restenosis was classified if a diameter reduction of 50% or more when compared with the original stent diameter was present.
Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) were calculated for evaluable stents, for stents with a diameter >3.0 mm, equal to 3.0 mm, and <3.0 mm as well as per patient (detection of at least one in-stent stenosis per patient, with unevaluable stents classified as having in-stent stenosis). Confidence intervals were calculated using the efficient-score method (corrected for continuity).28 As no patient had more than one lesion with in-stent stenosis, correction for in-patient correlation was not performed.
2 test was used to compare the rate of unevaluable stents depending on heart rate and stent type. Mean diameters of evaluable and unevaluable stents were compared using the MannWhitney U test. All tests were two-tailed and a P-value less than 0.05 was considered significant.
A sample size of 100 stented lesions was determined under the assumption of a restenosis rate of 20% in order to be able to prove a sensitivity of at least 75% of MDCT for detection of in-stent stenosis with a power of 75%.
| Results |
|---|
|
|
|---|
Eighty-four patients were initially screened for inclusion in the study. Twenty could not be included because of elevated serum creatinine level (>1.5 mg/dL, n=8), atrial fibrillation (n=7), previous stent-in-stent implantation (n=2), or because they refused participation (n=3), so that 64 patients underwent 64-slice MDCT.
Fifty-nine of all 102 stented lesions (58%) were classified as evaluable in MDCT. The mean diameter of stents classified as unevaluable was 3.03±0.31 mm (range: 2.54.0 mm), whereas the mean diameter of evaluable stents was 3.28±0.40 mm (range: 2.755.0 mm, P=0.0002). Of all 13 stents with a diameter of <3.0 mm, only one stent (2.75 mm diameter) was classified as evaluable (and correctly determined to be free of restenosis). Thirty-three of all 57 stents with a diameter of 3.0 mm and 25 of 32 stents with a diameter >3.0 mm were classified as evaluable. Motion artefacts were visually detectable in 15 stents, which were classified as unevaluable (Figure 2). Nineteen of 46 stents in patients with a heart rate
60 b.p.m. and 24 of 56 stents in patients with a heart rate >60 b.p.m. were classified as unevaluable (P=0.08).
|
Stent type had an influence on evaluability. Of all 31 stents with a diameter
3.0 mm that were classified as unevaluable, 30 were of the BxSonic® (n=20), or Cypher® type (n=10), both share the same strut system with a strut thickness of 0.14 mm. In comparison, only one of nine Taxus® stents, which have strut thickness of 0.13 mm, was classified as unevaluable (P=0.001 for Cypher in comparison with Taxus and 0.03 for BxSonic in comparison with Taxus). Overall, 12 out of 102 stented lesions showed significant in-stent stenosis in QCA. Six of these were correctly detected by MDCT (overall sensitivity 50%, confidence interval 2277%). Fifty-one of 90 stented lesions without restenosis were correctly classified by MDCT (overall specificity 57%, confidence interval 4667%).
In the 59 evaluable stented lesions, MDCT correctly detected six of seven in-stent restenoses (sensitivity 86%, confidence interval 4299%) and correctly ruled out the presence of in-stent restenosis in 51 of 52 cases (specificity 98%, confidence interval 8899%). PPV was 86% (confidence interval 4299%) and NPV was 98% (confidence interval 8899%) (Table 2; Figures 3 and 4).
|
|
|
| Discussion |
|---|
|
|
|---|
Owing to artefacts caused by metal, the visualization of the lumen within coronary artery stents by MDCT is more challenging than the assessment of the native coronary arteries. Although accuracies reported for the assessment of coronary artery stenoses in non-stented arteries by 16- and 64-slice MDCT are high, only a small number of studies have evaluated the value of MDCT to detect in-stent restenosis.2127 In those studies, most performed by 16-slice CT, the rate of evaluable stents was low and even in assessable stentswith the exception of large stents implanted in the left main coronary arterysensitivity for the detection of in-stent restenosis was only between 54 and 83%. Next to the type of scanner used, several factors that influence assessability of stents have been identified. They include the type of stent, as shown in phantom studies.29,30 Also, there have been indications that stent diameter plays a significant role. In the study by Schuijf et al.,22 28% of stents with a diameter
3.0 mm, but only 10% of stents with a diameter of >3.0 mm, were unevaluable. In a study by Gilard et al.,26 which included only stents implanted in the left main coronary arterywith a large mean diameter of 3.9 mmall stents were assessable and all four restenoses were detectable by CT. In another study by the same groupwith 232 included stents, the largest such study to date49% of stents with a diameter
3.0 mm and 18% of stents with a diameter >3.0 mm could be evaluated.21 Gaspar et al.25 found that with 40-slice CT (which has a similar spatial resolution to the 64-slice scanner), all of 111 stents were evaluable (with a mean stent diameter of 3.3 mm), but only 72% of in-stent restenoses were detected. Our series confirms that even with improved scanner technology, evaluation of implanted coronary artery stents remains challenging. Similar to the earlier studies performed by 16- and 40-slice MDCT, we also found a significant influence of stent diameter on evaluability, with 3.5 mm being a threshold below which the rate of evaluable stents is very low. Stent type was associated with evaluability because Taxus stents (with a strut thickness of 0.13 mm) were evaluable at a substantially higher rate than those with a strut thickness of 0.14 mm (Cypher and BxSonic). After exclusion of all unevaluable stents, sensitivity for the detection of in-stent restenosis was 86% (with a specificity of 98%), which indicates that clinical applications might be possible if the problem of stent evaluability was solved. However, the high frequency of cases with un-interpretable image qualityan overall rate of 42% in our series and 32% even in patients with stents of >3 mm diameterprecludes the application of MDCT for coronary angiography in unselected patients with implanted coronary artery stents.
Although we report on the first series of consecutive patients with implanted coronary artery stents evaluated by 64-slice CT to date, the study suffers several limitations. The overall rate of stents with a significant stenosis was low, with only 12 restenoses in 102 stented lesions and only seven restenoses in evaluable stents. The low rate of restenosis was below our assumptions for the power calculation and led to wide confidence intervals. The fact that 97% of included stents were of the type Taxus, Cypher, or BxSonic constitutes an advantage of our study, because it permits more systematic assessment of the influence of stent diameter, but also limits our results to only a small number of all coronary stents that are commercially available. Finally, we used a simple definition of in-stent restenosis (diameter reduction >50%), although many other potential definitions, such as per-segment restenosis and late lumen loss (absolute decrease of minimal stent lumen from post-intervention to follow-up), have also been proposed and there is no straightforward correlation between any definition of in-stent stenosis and its haemodynamic relevance.
In conclusion, our study helps to identify factors that influence the assessability of coronary artery stents by 64-slice CT, namely, stent type and diameter. It shows that under certain conditions, the detection of in-stent restenosis might be possible with an accuracy that could permit clinical applications, but that the relatively high rate of unevaluable stents currently does not allow to use MDCT for coronary angiography in unselected patients with implanted stents in coronary arteries or bypass grafts.
Conflict of interest: S.A. and A.K. have received research grants from Siemens Medical Solutions.
| References |
|---|
|
|
|---|
- Holmes DR Jr, Leon MB, Moses JW, Popma JJ, Cutlip D, Fitzgerald PJ, Brown C, Fischell T, Wong SC, Midei M, Snead D, Kuntz RE. (2004) Analysis of 1-year clinical outcomes in the SIRIUS trial: a randomized trial of a sirolimus-eluting stent versus a standard stent in patients at high risk for coronary restenosis. Circulation 109:634640.
[Abstract/Free Full Text] - Moses JW, Leon MB, Popma JJ, Fitzgerald PJ, Holmes DR, O'Shaughnessy C, Caputo RP, Kereiakes DJ, Williams DO, Teirstein PS, Jaeger JL, Kuntz RE. SIRIUS Investigators. (2003) Sirolimus-eluting stents versus standard stents in patients with stenosis in a native coronary artery. N Engl J Med 349:13151323.
[Abstract/Free Full Text] - Grube E, Silber S, Hauptmann KE, Mueller R, Buellesfeld L, Gerckens U, Russell ME. (2003) TAXUS I: six- and twelve-month results from a randomized, double-blind trial on a slow-release paclitaxel-eluting stent for de novo coronary lesions. Circulation 107:3842.
[Abstract/Free Full Text] - Morice MC, Colombo A, Meier B, Serruys P, Tamburino C, Guagliumi G, Sousa E, Stoll HP. REALITY Trial Investigators. (2006) Sirolimus- vs. paclitaxel-eluting stents in de novo coronary artery lesions: the REALITY trial: a randomized controlled trial. JAMA 295:895904.
[Abstract/Free Full Text] - Windecker S, Remondino A, Eberli FR, Juni P, Raber L, Wenaweser P, Togni M, Billinger M, Tuller D, Seiler C, Roffi M, Corti R, Sutsch G, Maier W, Luscher T, Hess OM, Egger M, Meier B. (2005) Sirolimus-eluting and paclitaxel-eluting stents for coronary revascularization. N Engl J Med 353:653662.
[Abstract/Free Full Text] - Mauri L, O'Malley AJ, Popma JJ, Moses JW, Leon MB, Holmes DR Jr, Teirstein PS, Cutlip DE, Donahoe D, Kuntz RE. (2005) Comparison of thrombosis and restenosis risk from stent length of sirolimus-eluting stents versus bare metal stents. Am J Cardiol 95:11401145.[CrossRef][Web of Science][Medline]
- Ardissino D, Cavallini C, Bramucci E, Indolfi C, Marzocchi A, Manari A, Angeloni G, Carosio G, Bonizzoni E, Colusso S, Repetto M, Merlini PA. SES-SMART Investigators. (2004) Sirolimus-eluting vs. uncoated stents for prevention of restenosis in small coronary arteries: a randomized trial. JAMA 292:27272734.
[Abstract/Free Full Text] - Ge L, Airoldi F, Iakovou I, Cosgrave J, Michev I, Sangiorgi GM, Montorfano M, Chieffo A, Carlino M, Corvaja N, Colombo A. (2005) Clinical and angiographic outcome after implantation of drug-eluting stents in bifurcation lesions with the crush stent technique: importance of final kissing balloon post-dilation. J Am Coll Cardiol 46:613620.
[Abstract/Free Full Text] - Zellweger MJ, Weinbacher M, Zutter AW, Jeger RV, Mueller-Brand J, Kaiser C, Buser PT, Pfisterer ME. (2003) Long-term outcome of patients with silent versus symptomatic ischemia six months after percutaneous coronary intervention and stenting. J Am Coll Cardiol 42:3340.
[Abstract/Free Full Text] - Mollet NR, Cademartiri F, Krestin GP, McFadden EP, Arampatzis CA, Serruys PW, de Feyter PJ. (2005) Improved diagnostic accuracy with 16-row multi-slice computed tomography coronary angiography. J Am Coll Cardiol 45:128132.
[Abstract/Free Full Text] - Morgan-Hughes GJ, Roobottom CA, Owens PE, Marshall AJ. (2005) Highly accurate coronary angiography with submillimetre, 16 slice computed tomography. Heart 91:308313.
[Abstract/Free Full Text] - Schuijf JD, Bax JJ, Salm LP, Jukema JW, Lanb HJ, van der Wall EE, de Roos A. (2005) Noninvasive coronary imaging and assessment of left ventricular function using 16-slice computed tomography. Am J Cardiol 95:571574.[CrossRef][Web of Science][Medline]
- Hoffmann MHK, Shi H, Schmitz BL, Schmid FT, Lieberknecht M, Schulze R, Ludwig B, Kroschel U, Jahnke N, Haerer W, Brambs HJ, Aschoff AJ. (2005) Noninvasive coronary angiography with multislice computed tomography. JAMA 293:24712478.
[Abstract/Free Full Text] - Achenbach S, Ropers D, Pohle FK, Raaz D, von Erffa J, Yilmaz A, Muschiol G, Daniel WG. (2005) Detection of coronary artery stenoses using multi-detector CT with 16x0.75 mm collimation and 375 ms rotation. Eur Heart J 26:19781986.
[Abstract/Free Full Text] - Leschka S, Alkadhi H, Plass A, Desbiolles L, Grünenfelder J, Marincek B, Wildermuth S. (2005) Accuracy of MSCT coronary angiography with 64-slice technology: first experience. Eur Heart J 26:14821487.
[Abstract/Free Full Text] - Raff GJ, Gallagher MJ, ÓNeill WW, Goldstein JA. (2005) Diagnostic accuracy of noninvasive angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46:552557.
[Abstract/Free Full Text] - Leber AW, Knez A, von Ziegler F, Becker A, Nikolaou K, Paul S, Wintersperger B, Reiser M, Becker CR, Steinbeck G, Boekstegers P. (2005) Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography. A comparative study with quantitative coronary angiography and intravascular ultrasound. J Am Coll Cardiol 46:147154.
[Abstract/Free Full Text] - Mollet NR, Cademartiri F, van Mieghem CA, Runza G, McFadden EP, Baks T, Serruys PW, Krestin GP, de Feyter PJ. (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112:23182323.
[Abstract/Free Full Text] - Ropers D, Rixe J, Anders K, Küttner A, Baum U, Bautz W, Daniel WG, Achenbach S. (2006) Usefulness of multidetector row computed tomography with 64x0.6 mm collimation and 330-ms rotation for the noninvasive detection of significant coronary artery stenoses. Am J Cardiol 97:343348.[CrossRef][Web of Science][Medline]
- Fine JJ, Hopkins CB, Ruff N, Newton FC. (2006) Comparison of accuracy of 64-slice cardiovascular computed tomography with coronary angiography in patients with suspected coronary artery disease. Am J Cardiol 97:173174.[CrossRef][Web of Science][Medline]
- Gilard M, Cornily JC, Pennec PY, Le Gal G, Nonent M, Mansourati J, Blanc JJ, Boschat J. (2006) Assessment of coronary artery stents by 16 slice computed tomography. Heart 92:5861.
[Abstract/Free Full Text] - Schuijf JD, Bax JJ, Jukema JW, Lamb HJ, Warda HM, Vliegen HW, de Roos A, van der Wall EE. (2004) Feasibility of assessment of coronary stent patency using 16-slice computed tomography. Am J Cardiol 94:427430.[CrossRef][Web of Science][Medline]
- Kitagawa T, Fujii T, Tomohiro Y, Maeda K, Kobayashi M, Kunita E, Sekiguchi Y. (2006) Noninvasive assessment of coronary stents in patients by 16-slice computed tomography. Int J Cardiol 109:188194.[CrossRef][Web of Science][Medline]
- Cademartiri F, Mollet N, Lemos PA, Pugliese F, Baks T, McFadden EP, Krestin GP, de Feyter PJ. (2005) Usefulness of multislice computed tomographic coronary angiography to assess in-stent restenosis. Am J Cardiol 96:799802.[CrossRef][Web of Science][Medline]
- Gaspar T, Halon DA, Lewis BS, Adawi S, Schliamser JE, Rubinshtein R, Flugelman MY, Peled N. (2005) Diagnosis of coronary in-stent restenosis with multidetector row spiral computed tomography. J Am Coll Cardiol 46:15731579.
[Abstract/Free Full Text] - Gilard M, Cornily JC, Rioufol G, Finet G, Pennec PY, Mansourati J, Blanc JJ, Boschat J. (2005) Noninvasive assessment of left main coronary stent patency with 16-slice computed tomography. Am J Cardiol 95:110112.[CrossRef][Web of Science][Medline]
- Van Mieghem CA, Cademartiri F, Mollet NR, Malagutti P, Valgimigli M, Meijboom WB, Pugliese F, McFadden EP, Ligthart J, Runza G, Bruining N, Smits PC, Regar E, van der Giessen WJ, Sianos G, van Domburg R, de Jaegere P, Krestin GP, Serruys PW, de Feyter PJ. (2006) Multislice spiral computed tomography for the evaluation of stent patency after left main coronary artery stenting: a comparison with conventional coronary angiography and intravascular ultrasound. Circulation 114:645653.
- Newcombe RG. (1998) Two-sided confidence intervals for the single proportion: comparison of seven methods. Stat Med 17:857872.[CrossRef][Web of Science][Medline]
- Mahnken AH, Muhlenbruch G, Seyfarth T, Flohr T, Stanzel S, Wildberger JE, Gunther RW, Kuettner A. (2006) 64-slice computed tomography assessment of coronary artery stents: a phantom study. Acta Radiol 47:3642.[CrossRef][Web of Science][Medline]
- Maintz D, Seifarth H, Raupach R, Flohr T, Rink M, Sommer T, Ozgun M, Heindel W, Fischbach R. (2006) 64-slice multidetector coronary CT angiography: in vitro evaluation of 68 different stents. Eur Radiol 16:818826.[CrossRef][Web of Science][Medline]
This article has been cited by other articles:
![]() |
F. Wolf, F. Cademartiri, C. Loewe, A. Stadler, M. Weber, J. Lammer, and G. M. Feuchtner Evaluation of Coronary Stents With 64-MDCT: In Vitro Comparison of Scanners From Four Vendors Am. J. Roentgenol., September 1, 2009; 193(3): 787 - 794. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. J. de Feyter, S. Achenbach, and K. Nieman CHAPTER 6 Cardiovascular Computed Tomography ESC Textbook of Cardiovascular Medicine, January 1, 2009; 2(1): med-9780199566990-chapter - med-9780199566990-chapter. [Abstract] [Full Text] [PDF] |
||||
![]() |
G Mowatt, J A Cook, G S Hillis, S Walker, C Fraser, X Jia, and N Waugh 64-Slice computed tomography angiography in the diagnosis and assessment of coronary artery disease: systematic review and meta-analysis Heart, November 1, 2008; 94(11): 1386 - 1393. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. A. Bluemke, S. Achenbach, M. Budoff, T. C. Gerber, B. Gersh, L. D. Hillis, W. G. Hundley, W. J. Manning, B. F. Printz, M. Stuber, et al. Noninvasive Coronary Artery Imaging: Magnetic Resonance Angiography and Multidetector Computed Tomography Angiography: A Scientific Statement From the American Heart Association Committee on Cardiovascular Imaging and Intervention of the Council on Cardiovascular Radiology and Intervention, and the Councils on Clinical Cardiology and Cardiovascular Disease in the Young Circulation, July 29, 2008; 118(5): 586 - 606. [Full Text] [PDF] |
||||
![]() |
F Pugliese, A C Weustink, C Van Mieghem, F Alberghina, M Otsuka, W B Meijboom, N van Pelt, N R Mollet, F Cademartiri, G P Krestin, et al. Dual source coronary computed tomography angiography for detecting in-stent restenosis Heart, July 1, 2008; 94(7): 848 - 854. [Abstract] [Full Text] [PDF] |
||||
![]() |
W T Roberts, J J Bax, and L C Davies Cardiac CT and CT coronary angiography: technology and application Heart, June 1, 2008; 94(6): 781 - 792. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. Eichhorn, C. Jourdan, S. L. Hill, S. V. Raman, J. P. Cheatham, and F. R. Long CT of Pediatric Vascular Stents Used to Treat Congenital Heart Disease Am. J. Roentgenol., May 1, 2008; 190(5): 1241 - 1246. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Schroeder, S. Achenbach, F. Bengel, C. Burgstahler, F. Cademartiri, P. de Feyter, R. George, P. Kaufmann, A. F. Kopp, J. Knuuti, et al. Cardiac computed tomography: indications, applications, limitations, and training requirements: Report of a Writing Group deployed by the Working Group Nuclear Cardiology and Cardiac CT of the European Society of Cardiology and the European Council of Nuclear Cardiology Eur. Heart J., February 2, 2008; 29(4): 531 - 556. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Bateman Business aspects of cardiovascular computed tomography: tackling the challenges. J. Am. Coll. Cardiol. Img., January 1, 2008; 1(1): 111 - 118. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Abdulla, S. Z. Abildstrom, O. Gotzsche, E. Christensen, L. Kober, and C. Torp-Pedersen 64-multislice detector computed tomography coronary angiography as potential alternative to conventional coronary angiography: a systematic review and meta-analysis Eur. Heart J., December 2, 2007; 28(24): 3042 - 3050. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Schuijf, G. Pundziute, J. W. Jukema, H. J. Lamb, J. C. Tuinenburg, B. L. van der Hoeven, A. de Roos, J. H. C. Reiber, E. E. van der Wall, M. J. Schalij, et al. Evaluation of Patients with Previous Coronary Stent Implantation with 64-Section CT Radiology, November 1, 2007; 245(2): 416 - 423. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Cademartiri, J. D. Schuijf, F. Pugliese, N. R. Mollet, J. W. Jukema, E. Maffei, L. J. Kroft, A. Palumbo, D. Ardissino, P. W. Serruys, et al. Usefulness of 64-Slice Multislice Computed Tomography Coronary Angiography to Assess In-Stent Restenosis J. Am. Coll. Cardiol., June 5, 2007; 49(22): 2204 - 2210. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||