Skip Navigation


European Heart Journal Advance Access originally published online on October 2, 2006
European Heart Journal 2006 27(24):3033-3038; doi:10.1093/eurheartj/ehl273
This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
27/24/3033    most recent
ehl273v1
Right arrow E-letters: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when E-letters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Related articles in EHJ
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (14)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Bouvier, E.
Right arrow Articles by Cohen-Solal, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bouvier, E.
Right arrow Articles by Cohen-Solal, A.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?

© The European Society of Cardiology 2006. All rights reserved. For Permissions, please e-mail: journals.permissions@oxfordjournals.org

Diagnosis of aortic valvular stenosis by multislice cardiac computed tomography

Erik Bouvier1,2,*, Damien Logeart2, Jean-Louis Sablayrolles1, Jacques Feignoux1, Claude Scheublé1, Thierry Touche1, Gabriel Thabut3 and Alain Cohen-Solal2

1 Service de Radiologie et de Cardiologie, Centre Cardiologique du Nord, Saint Denis, France
2 Service de Cardiologie, Hôpital Lariboisière, Paris, France
3 Service de Pneumologie, Hôpital Bichat, Paris, France

Received 3 January 2006; revised 5 July 2006; accepted 11 September 2006; online publish-ahead-of-print 2 October 2006.

* Corresponding author. Tel: +33 1 40 88 60 00; fax: +33 1 40 88 62 00. E-mail address: erik.bouvier{at}wanadoo.fr

See page 2923 for the editorial comment on this article (doi:10.1093/eurheartj/ehl375)


    Abstract
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusion
 References
 
Aims Current improvements in spatial, temporal, and contrast resolution of multislice computed tomography (CT) could be useful in the assessment of valvular diseases. We evaluated the diagnostic accuracy of multislice CT for the identification and quantification of aortic valvular stenosis (AS), compared with echography.

Methods and results Consecutive patients, referred for coronary CT, were evaluated for AS, by the use of standard electrocardiography-gated 16-slice CT protocol. Multiplanar reformat was applied to systolic phases of the cardiac cycle, with projection on thick slices for measuring the aortic valvular area (AVA). CT results were compared with echocardiographic-based measurement of the AVA. Among 107 enrolled patients, CT analysis of the AVA was feasible in 103. Among the 30 patients with AS, Bland–Altman analysis showed good agreement between the two methods [mean difference –7 mm2 (–40–25 mm2)].

Conclusion CT analysis of aortic valve is feasible in most cases and allows for reliable diagnosis and quantification of AS.

Key Words: Aortic stenosis • Diagnosis • Cardiac computed tomography


    Introduction
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusion
 References
 
Aortic stenosis (AS) is the most frequent valvular disease in Western countries, and the incidence of degenerative AS, the main cause of AS, should still increase because of the ever-increasing life expectancy. The timing for surgical treatment is the main concern in medical management of AS and requires accurate measurement of the aortic valvular area (AVA).1

Catheterism has been the standard method of evaluation for several decades but is declining in use because of concerns regarding the risk of catheter-related damage such as stroke2 and technical limitations. Today, Doppler echocardiography is the current reference technique1,3 providing insights into both aortic valve as well as left ventricular (LV) function. The Doppler-derived continuity equation has been largely validated, but entails technical limitations such as suboptimal apical acoustic windows, important angulation of the ascending aorta with the LV outflow tract, difficult measurement of the LV outflow tract diameter, and associated subaortic stenosis or aortic regurgitation. These limitations result in relatively high inter- and intra-observer variability.4,5 In addition, the haemodynamic model underlying the Doppler measurement fails in cases of low cardiac output.6

Meanwhile, recent technical improvements in multislice electrocardiography (ECG)-gated computerized tomography (MSCT) allows for visualization of the aortic valve throughout the cardiac cycle. In addition, cardiac MSCT gives information about the left ventricle and coronary artery anatomy. Thus, cardiac MSCT could be an alternative and integrated imaging technique in the management of valvular disease. However, MSCT has not been evaluated for AS.

Our aim was to compare 16-slice MSCT with Doppler echocardiography for quantifying AVA and diagnosing AS.


    Methods
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusion
 References
 
Population
From September 2003 to February 2005, we prospectively enrolled 131 consecutive patients referred for cardiac CT (LightSpeed Pro16 by GE Healthcare, Waukesha, WI, USA). Patients were referred for coronary artery imaging because of chest pain, abnormal ascending aorta, mediastinal/pericardial abnormalities in a peri-operative setting, or congenital heart disease. Exclusion criteria were renal failure (serum creatinine >150 µmol/L), severe allergy to iodine contrast, atrial fibrillation with fast ventricular rhythm, or sinus tachycardia >100/min. The study complied with the declaration of Helsinki and the local Ethics Committee approved the research protocol.

MSCT method
CT was acquired in a breath-hold, eased by oxygen supply, and ECG-gated, after verification of proper R-waves. A reference slide acquired at the level of the ascending aorta, regularly updated, allowed for detecting the contrast bolus and triggering the acquisition, to avoid the use of test contrast bolus, to shorten the examination time, and to optimize the synchronization between injection and acquisition.

Standard cardiac acquisition parameters were 16 slices, SFOV 250 mm, slice thickness 0.625 mm, reconstruction interval 0.625, 120 kV, 600 mA, 220–250 mA s, and scan rotation 0.4 s. No ECG mA-modulation was used. Rotation speed and pitch were automatically adjusted to the acquisition protocols and the heart rate. Acquisition time varied between 18 and 25 s, depending on anatomical coverage and pitch. The temporal resolution depended on rotation time and reconstruction mode and ranged from 200 to 80 ms. Contrast media (320–350 mg I/ml) was infused in triple bolus through a 18-gauge catheter in an antebrachial vein with use of a dual-barrel injector (90 cc of contrast media at 5 cc/s, 30 cc of contrast medium diluted with 30 cc of isotonic solution at 3.5 cc/s, washed out by 40 cc of isotonic solution at 3.5 cc/s). The reconstruction algorithm depended on heart rate: segment mode under 65 b.p.m. (half rotation), burst mode (90° sector on two consecutive beats) between 65 and 85 b.p.m., and burst plus mode (45° sector on four consecutive beats) over 85 b.p.m. Phases from 0 to 90% every 10% were systematically reconstructed to allow for imaging both coronary arteries, valvular apparatus, and cardiac chambers.

CT data were analysed by the use of Advantage Workstation (GE Healthcare, Version 4.2). CT readers (E.B. and J.L.S.) were blinded to the echographic results. Multiplanar reformat was applied to systolic phases of the cardiac cycle (from 0 to 40% of the cardiac cycle). Imaging planes for planimetry were chosen from the sagital and coronal views to include the orifice of the aortic valve. Slice thickness was then increased to >5 mm to ensure inclusion of the whole aortic valve orifice. Starting from the standard cardiac preset, window/centre setting was adjusted to minimize the blooming artefact around valvular calcifications. Use of both maximal intensity projection (MIP) and minimal intensity projection (MinIP) allowed for delineation of the AVA on each systolic phase (Figures 1 and 2). Finally, the AVA was the maximum value of systolic cusps opening. Phases displaying cusps blurred by an opening or closing motion were excluded.


Figure 2731
View larger version (161K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 1 Representative CT of severe aortic stenosis. CT imaging shows fibrous cusps with few calcifications; AVA is 64 mm2.

 

Figure 2732
View larger version (95K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 2 Representative CT of moderate aortic stenosis, with calcified fusion of the anterior commissure, AVA 119 mm2. MinIP rendering (left) improves depiction of low densities, like soft tissues. MIP rendering (right) enhances high densities, like calcifications. The planimetry is first performed on the MinIP rendering, then the MIP rendering allows to exclude calcifications from the planimetry, if necessary.

 
Echocardiography
Echocardiographic examination was systematically performed by a trained operator (C.S. or T.T.) blinded to the CT results. The mean delay between CT and echocardiographic examination was 2 (1–13) days. The AVA was measured by Doppler-based continuity equation and/or 2D quantification from transesophageal echocardiography if necessary. Patients with a normal M-mode aortic valve (valvular opening >15 mm) and normal transvalvular velocity (peak systolic Doppler velocity <1.5 m/s) were classified as non-stenotic. Otherwise, the valvular area was calculated with the continuity equation, SAO=({pi}xD2/4)xVTILVOT/VTIAO, where D is the LVOT diameter and VTI the Doppler-based velocity–time integrals through the aortic valve (VTIAO) or LVOT (VTILVOT). We averaged VTIAO on the three highest velocities and VTILVOT on three consecutive heartbeats when a Doppler click of aortic valve closure was identified. Improper imaging of the aortic valve and ascending aorta, mainly in the case of limited apical acoustic window, required transesophageal echocardiography to allow direct planimetry of the AVA. Transesophageal echocardiography was performed with the use of a multiplane probe. The planimetry was usually performed in a 30–50° plane, when the aortic valve annulus appeared circular and all cusps were seen at the same time. Subtle movements of the probe allowed for precise identification of the smallest AVA at the time of maximal systolic opening. Three measures were averaged. Stenosis was graded according to the American College of Cardiology (ACC)/American Heart Association (AHA) classification: mild (150–200 mm2), moderate (>100–150 mm2), severe (>75–100 mm2), and critical (≤75 mm2).1 LV ejection fraction was measured by the use of the Simpson biplane method.

Statistical analysis
Given the uncertainty regarding the expected agreement between the two techniques (echography and MSCT) and the exploratory nature of such a pilot study, 30 patients with significant AS were empirically included. Continuous variables are reported as mean and standard deviation or median and range, and compared using t-test or Mann–Withney U test. Categorical variables are reported as count and percentages and compared using {chi}2 tests. The agreement between CT and echography was displayed using the Bland and Altman plots.7 Among CT examinations, 30 cases were randomly and independently analysed by two trained operators who repeated the whole analysis of CT from the same acquired data in order to evaluate inter- and intra-observer variability regarding the planimetry of the AVA; the intraclass correlation coefficient was used for this analysis. Confidence intervals were by bootstrap resampling. All tests were two-sided with P<0.05 indicating statistical significance. Tests were not corrected for multiple testing.


    Results
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusion
 References
 
Among 131 patients initially assessed for inclusion, five refused to give their consent, eight were excluded because of renal failure or allergy, seven because of heart rate >100 b.p.m., and four had no reliable Doppler measurement and refused TEE. Finally, CT analysis of the aortic area was not feasible in four patients because of a bad signal/noise ratio or arrhythmia-related artefacts. Table 1 describes the characteristics of the 103 included patients. Echocardiographic LV ejection fraction was 61±14%. Echocardiographic AS was diagnosed in 30 patients (mean AVA=80±27 mm2) and was graded as follows: 15 critical, 11 severe, four moderate or mild. Transesophageal imaging was required in 17 patients (only three with AS).


View this table:
[in this window]
[in a new window]

 
Table 1 Main characteristics of patients

 
During the CT examination, the mean heart rate was 71±16 b.p.m. and the optimal phase for aortic valvular measurement was 10% of the R–R cycle in 9 patients, 20% in 67 patients, 30% in 30 patients and 40% in one patient. The mean CT-based AVA was 244±133 mm2 (range 27–555 mm2). CT imaging identified 30 instances of AS, graded as 12 critical, 12 severe, four moderate, and two mild. The intra- and inter-observer agreements were very good with intraclass correlation coefficient of 0.99 for both (95% CI 0.99–1.00). Heavily calcified valves were easier to delineate with the MIP technique. Purely fibrous stenoses were analysed with the MinIP technique. Mixed stenoses required concordant analysis of both rendering techniques. Bicuspid was diagnosed in six patients from both CT and echocardiographic analysis. The acquisition time of CT data ranged between 18 and 25 s, the examination time between 7 and 11 min, and the post-processing time for AVA planimetry ranged between 10 and 18 min.

Among the 30 patients with echocardiographic AS and AVA≤200 mm2, the CT-based AVA was 87±33 mm2. The Bland and Altman analysis showed good agreement between echography and MSCT (Figure 3) with a mean difference of –7 mm2 (95% CI: –40–25 mm2).


Figure 2733
View larger version (12K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 3 Bland and Altman analysis of aortic stenoses (≤200 mm2).

 
According to the ACC/AHA classification, CT and echocardiographic grading matched in 21 patients and differed by one class in nine patients: seven were overestimated and two were underestimated by CT imaging (Table 2).


View this table:
[in this window]
[in a new window]

 
Table 2 ACC/AHA grading of aortic stenosis according to cardiac CT and echocardiography

 

    Discussion
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusion
 References
 
This study shows that CT imaging of AS is feasible and reliable when compared with echocardiography. We demonstrate the relevance of 16-slice MSCT for diagnosing and grading AS.

Cardiac CT is a fast, non-invasive, well-tolerated, and reproducible imaging method with reliable diagnostic accuracy when compared with echocardiography. CT-based and echocardiographic measurements of the AVA are closely related and the ACC/AHA classification is similar for the two techniques, which should lead to similar therapeutic management.

Advantages and recent improvements in MSCT analysis
The spatial resolution of cardiac CT has improved since the advent of spiral 16-slice acquisitions: voxel size is now reduced to 0.625 mm in the Z-axis and 0.5 mm in the X and Y-axes, which compares favourably with echocardiography measurement. Temporal resolution has improved with increased rotating speed; our 16-slice CT achieves a complete revolution in 400 ms and, thus, allows for a temporal resolution of 80–200 ms. In the 60–100 b.p.m. range, aortic ejection accounts for about 30% of the cardiac cycle (i.e. 180–300 ms of the aortic valve opening, which can be slightly reduced in critical AS). Interestingly, the newer 64-slice CT will further improve the temporal resolution. In addition, cardiac CT results in a unique 3D image, which allows for analysis of the aortic valve in any plane and with various rendering techniques (multiplanar reformat, volume rendering, navigation) (Figure 4). Radiography is less sensitive to calcium artefacts than ultrasonography and thus more reliable when calcifications are present. Importantly, intra- and inter-observer variabilities of CT measurements were very low in our study. In contrast to Doppler-based measurement, the pure anatomic approach of CT imaging is independent of inotropism as well as load, which may explain some differences between the two techniques.810 In our study, CT-based AVA is slightly larger than Doppler-based AVA, the latter representing the functional, effective valve opening which is smaller than the anatomic valve. Cardiac CT displays anatomic information comparable to transesophageal echocardiography11 (shape of the orifice, commissure, bicuspid valve, AVA, localization of coronary ostia, grading of calcification, enlargement of the aorta, atheroma, and calcification of the ascending aorta) but also offers imaging of the coronary tree and the LV function in the same examination. Finally, the need for preoperative coronary angiography could be reduced because MSCT is becoming a reliable tool for diagnosing coronary artery disease.12,13


Figure 2734
View larger version (164K):
[in this window]
[in a new window]
[Download PowerPoint slide]
 
Figure 4 Bicuspid aortic valve, diastole (left) and systole (right), multiplanar reformat (top), and 3D navigation rendering (bottom).

 
Technical limits of MSCT
The heart rate variability remains the main limiting factor of CT analysis; atrial fibrillation, unstable sinus rhythm, and ventricular premature beats are responsible for band artefacts (two patients in our study). At the time of this study, we did not use beta-blockers, but now such medications are frequently prescribed and result in less heart rate-related artefacts. In addition, patients must be able to hold a breath during 25–30 s with 16-slice MSCT, but this duration is now reduced to 10–15 s with newer 64-slice MSCT. Heavily calcified cusps are a theoretical limitation for CT analysis because of the blooming artefact. In fact, such an effect is limited to adjacent voxels after proper adjustment of window/centre setting and is less worrying than with coronary imaging. In patients with low cardiac output, echography using dobutamine stress is advised because of the risk of AVA underestimation, and ultrasound analysis appears more suitable for such cases than MSCT. It should be reminded that MSCT use is limited in patients with renal failure and allergy and requires caution in patients with severe LV dysfunction. At last, CT-related radiation exposition is close to cardiac catheterism;14 the delivery dose from CT ranged between 10 and 17 mSv in our study.

Comparison with magnetic resonance imaging
Several authors demonstrated that MRI could be used to analyse the AVA.1518 However, when compared with MSCT, MRI is not feasible in patients with pacemakers, implantable defibrillators, vascular clips or claustrophobia, it does not allow for visualization of calcifications, signal voids are caused by turbulent flows, it offers poor imaging of the coronary tree, and it is time-consuming (40–60 min) and more expensive.

Limitations of the study
The sample size of the study was relatively small. Patients with mild/moderate AS are under-represented because of recruitment bias, but the strong diagnostic accuracy and the relationship we showed in patients with severe AS suggest that MSCT should be reliable in all ranges of valve areas. In our study, we did not have the advantage of a gold standard such as haemodynamics or surgical findings. CT measurements were compared only with echocardiographic measurements. In fact, this study aimed to show how CT results are close to echocardiographic results and how CT examination could possibly substitute for the latter.


    Conclusion
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusion
 References
 
Cardiac CT is a promising new technique for comprehensive assessment of aortic valvular anatomy and for diagnosing AS. Although ultrasound examination is the current first-line imaging, MSCT can be of value when echocardiographic examination is technically difficult and/or does not match the clinical data. The combined use of echocardiography and CT imaging challenges the requirement for cardiac catheterism, especially in cases of pathological ascending aorta or low probability of coronary disease.

Conflict of interest: J.-L.S. has received fees for speaking a meetings funded by GE Healthcare. All other authors have no conflict to declare.


    References
 Top
 Abstract
 Introduction
 Methods
 Results
 Discussion
 Conclusion
 References
 

  1. Bonow RO, Carabello B, de Leon AC Jr, Edmunds LH Jr, Fedderly BJ, Freed MD, Gaasch WH, McKay CR, Nishimura RA, O'Gara PT, O'Rourke RA, Rahimtoola SH. (1998) ACC/AHA guidelines for the management of patients with valvular heart disease: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee on Management of Patients with Valvular Heart Disease). J Am Coll Cardiol 32:1486–1588.[Free Full Text]
  2. Omran H, Schmidt H, Hackenbroch M, Ilien S, Bernhardt P, von der Recke G, Fimmers R, Flacke S, Layer G, Pohl C, Luderitz B, Schild H, Sommer T. (2003) Silent and apparent cerebral embolism after retrograde catheterisation of the aortic valve in valvular stenosis: a prospective, randomised study. Lancet 361:1241–1246.[CrossRef][Web of Science][Medline]
  3. Roger VL, Tajik AJ, Reeder GS, Hayes SN, Mullany CJ, Bailey KR, Seward JB. (1996) Effect of Doppler echocardiography on utilization of hemodynamic cardiac catheterization in the preoperative evaluation of aortic stenosis. Mayo Clin Proc 71:141–149.[Abstract]
  4. Geibel A, Gornandt L, Kasper W, Bubenheimer P. (1991) Reproducibility of Doppler echocardiographic quantification of aortic and mitral valve stenoses: comparison between two echocardiography centers. Am J Cardiol 65:1013–1021.
  5. Myreng Y, Molstad P, Endresen K, Ihlen H. (1990) Reproducibility of echocardiographic estimates of the area of stenosed aortic valves using the continuity equation. Int J Cardiol 26:349–354.[CrossRef][Web of Science][Medline]
  6. Burwash IG, Thomas DD, Sadahiro M, Pearlman AS, Verrier ED, Thomas R, Kraft CD, Otto CM. (1994) Dependence of Gorlin formula and continuity equation valve areas on transvalvular volume flow rate in valvular aortic stenosis. Circulation 89:827–835.
  7. Bland JM and Altman DG. (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310.[CrossRef][Web of Science][Medline]
  8. Baumgartner H, Kratzer H, Helmreich G, Kuehn P. (1990) Determination of aortic valve area by Doppler echocardiography using the continuity equation: a critical evaluation. Cardiology 77:101–111.[Web of Science][Medline]
  9. Rask LP, Karp KH, Eriksson NP. (1996) Flow dependence of the aortic valve area in patients with aortic stenosis: assessment by application of the continuity equation. J Am Soc Echocardiogr 9:295–299.[CrossRef][Medline]
  10. Cape EG, Jones M, Yamada I, VanAuker MD, Valdes-Cruz LM. (1996) Turbulent/viscous interactions control Doppler/catheter pressure discrepancies in aortic stenosis. The role of the Reynolds number. Circulation 94:2975–2981.
  11. Hoffmann R, Flchskampf FA, Hanrath P. (1993) Planimetry of orifice area in aortic stenosis using multiplane transesophageal echocardiography. J Am Coll Cardiol 22:529–534.[Abstract]
  12. Mollet NR, Cademartiri F, van Mieghem CA, Runza G, McFadden EP, Baks T, Serruys PW, Krestin GP, de Feyter PJ. (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112:2318–2323.
  13. Leber AW, Knez A, von Ziegler F, Becker A, Nikolaou K, Paul S, Wintersperger B, Reiser M, Becker CR, Steinbeck G, Boekstegers P. (2005) Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography: a comparative study with quantitative coronary angiography and intravascular ultrasound. J Am Coll Cardiol 46:147–154.[Abstract/Free Full Text]
  14. Stisova V. (2004) Effective dose to patient during cardiac interventional procedures. Radiat Prot Dosimetry 111:271–274.[Abstract/Free Full Text]
  15. de Roos A, Reichek N, Axel L, Kressel HY. (1989) Cine MR imaging in aortic stenosis. J Comput Assist Tomogr 13:421–425.[Web of Science][Medline]
  16. Friedrich MG, Schulz-Menger J, Poetsch T, Pilz B, Uhlich F, Dietz R. (2002) Quantification of valvular aortic stenosis by magnetic resonance imaging. Am Heart J 144:329–334.[CrossRef][Web of Science][Medline]
  17. John AS, Dill T, Brandt RR, Rau M, Ricken W, Bachmann G, Hamm CW. (2003) Magnetic resonance to assess the aortic valve area in aortic stenosis. J Am Coll Cardiol 42:519–526.[Abstract/Free Full Text]
  18. Kupfahl C, Honold M, Meinhardt G, Vogelsberg H, Wagner A, Mahrholdt H, Sechtem U. (2004) Evaluation of aortic stenosis by cardiovascular magnetic resonance imaging: comparison with established routine clinical techniques. Heart 90:893–901.[Abstract/Free Full Text]

Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us    What's this?

Related articles in EHJ:

Is there a role for multislice computed tomography in aortic stenosis?
Helmut Bamgartner
EHJ 2006 27: 2923-2924. [Extract] [FREE Full Text]  



This article has been cited by other articles:


Home page
Am. J. Roentgenol.Home page
E. J. Halpern, R. Mallya, M. Sewell, M. Shulman, and D. R. Zwas
Differences in Aortic Valve Area Measured with CT Planimetry and Echocardiography (Continuity Equation) Are Related to Divergent Estimates of Left Ventricular Outflow Tract Area
Am. J. Roentgenol., June 1, 2009; 192(6): 1668 - 1673.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Roentgenol.Home page
T. M. LaBounty, B. Sundaram, P. Agarwal, W. A. Armstrong, E. A. Kazerooni, and E. Yamada
Aortic Valve Area on 64-MDCT Correlates with Transesophageal Echocardiography in Aortic Stenosis
Am. J. Roentgenol., December 1, 2008; 191(6): 1652 - 1658.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart JHome page
K. K. Poh, R. A. Levine, J. Solis, L. Shen, M. Flaherty, Y.-J. Kang, J. L. Guerrero, and J. Hung
Assessing aortic valve area in aortic stenosis by continuity equation: a novel approach using real-time three-dimensional echocardiography
Eur. Heart J., October 2, 2008; 29(20): 2526 - 2535.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Roentgenol.Home page
G. M. Feuchtner, W. Dichtl, S. Muller, D. Jodocy, T. Schachner, A. Klauser, and J. O. Bonatti
64-MDCT for Diagnosis of Aortic Regurgitation in Patients Referred to CT Coronary Angiography
Am. J. Roentgenol., July 1, 2008; 191(1): W1 - W7.
[Abstract] [Full Text] [PDF]


Home page
HeartHome page
W T Roberts, J J Bax, and L C Davies
Cardiac CT and CT coronary angiography: technology and application
Heart, June 1, 2008; 94(6): 781 - 792.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Roentgenol.Home page
R. Ryan, S. Abbara, R. R. Colen, S. Arnous, M. Quinn, R. C. Cury, and J. D. Dodd
Cardiac Valve Disease: Spectrum of Findings on Cardiac 64-MDCT
Am. J. Roentgenol., May 1, 2008; 190(5): W294 - W303.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll Cardiol ImgHome page
L. F. Tops, S. C. Krishnan, J. D. Schuijf, M. J. Schalij, and J. J. Bax
Noncoronary applications of cardiac multidetector row computed tomography.
J. Am. Coll. Cardiol. Img., January 1, 2008; 1(1): 94 - 106.
[Abstract] [Full Text] [PDF]


Home page
HeartHome page
J.-P. Laissy, D. Messika-Zeitoun, J.-M. Serfaty, V. Sebban, E. Schouman-Claeys, B. Iung, and A. Vahanian
Comprehensive evaluation of preoperative patients with aortic valve stenosis: usefulness of cardiac multidetector computed tomography
Heart, September 1, 2007; 93(9): 1121 - 1125.
[Abstract] [Full Text] [PDF]


Home page
Eur Heart JHome page
H. Bamgartner
Is there a role for multislice computed tomography in aortic stenosis?
Eur. Heart J., December 2, 2006; 27(24): 2923 - 2924.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow FREE Full Text (PDF) Freely available
Right arrow All Versions of this Article:
27/24/3033    most recent
ehl273v1
Right arrow E-letters: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when E-letters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Related articles in EHJ
Right arrow Similar articles in this journal
Right arrow Similar articles in ISI Web of Science
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My Personal Archive
Right arrow Download to citation manager
Right arrow Search for citing articles in:
ISI Web of Science (14)
Right arrowRequest Permissions
Right arrow Disclaimer
Google Scholar
Right arrow Articles by Bouvier, E.
Right arrow Articles by Cohen-Solal, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Bouvier, E.
Right arrow Articles by Cohen-Solal, A.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us  
What's this?