Copyright © 2001 by the European Society of Cardiology.
In-vivo testing of digital cellular telephones in patients with implantable cardioverter-defibrillators
Cardiology Division, Patras University Medical School, Rio, Patras, Greece
revised October 24, 2000; accepted November 1, 2000
Abstract
Aims To investigate the susceptibility of implantable cardioverter defibrillators to electromagnetic interference generated by digital cellular telephones, functioning in both international transmission technologies: the Global System for Mobile Communication (GSM) and the Digital Cellular System (DCS 1800).
Methods and Results In 36 patients with transvenous implantable cardioverter defibrillators from two manufacturers (Medtronic and Guidant/CPI), cellular telephones with different levels of minimal and maximal power output were tested in the transmitting and receiving mode. Evaluation was performed in activated implantable defibrillators during spontaneous cardiac activity and continuous VVI or DDD pacing to assess possible electromagnetic interference. In two patients, appropriateness of ventricular fibrillation detection and therapy was judged during telephone testing. There was no damage, reprogramming, inappropriate shock therapy or pacing inhibition during the tests. In seven pre-pectoral Medtronic implantable defibrillators, transient electromagnetic interference caused 19 erroneous sensing events, when the operating phone was held in close vicinity to the programmer head. These pseudo-oversensing events, which did not result in logging of arrhythmia episodes in the device counter, were interpreted as an adverse interaction between the telephone and the programming device.
Conclusion Digital cellular telephones do not represent a risk to Medtronic and Guidant/CPI recipients of the specific implantable defibrillator models herein tested.
Key Words: Implantable cardioverter-defibrillators, electromagnetic interference, digital cellular telephones
f1 Correspondence: Antonis S. Manolis, MD, 41 Kourempana Street, Agios Dimitrios, Athens 17343, Greece.
References
- The Antiarrhythmics Versus Implantable Defibrillators (AVID) Investigators, A comparison of antiarrhythmic drug therapy with implantable defibrillators in patients resuscitated from near fatal ventricular arrhythmias. N Engl J Med 1997, 337, 1576, 83
- Moss AJ, Hall WJ, Cannom DS. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. Multicenter Automatic Defibrillator Implantation Trial Investigators. N Engl J Med. 1996;335:19331940
[Abstract/Free Full Text] - Buxton AE, Lee KL, Fisher JD, Josephson ME, Prystowsky EN, Hafley G. A randomized study of the prevention of sudden death in patients with coronary artery disease. Multicenter Unsustained Tachycardia Trial Investigators. N Engl J Med. 1999;34:18821890
- Kuck KH, Cappato R, Siebels J, Ruppel R. Randomized comparison of antiarrhythmic drug therapy with implantable defibrillators in patients resuscitated from cardiac arrest: the Cardiac Arrest Study Hamburg (CASH). Circulation. 2000;102:748754
[Abstract/Free Full Text] - Marchlinski FE, Flores BT, Buxton AE. The automatic implantable cardioverter defibrillator. Efficacy, complications and device failures. Ann Intern Med. 1986;104:481488[ISI][Medline]
- Haynes DL, Carillo RG, Findlay G, Embrey M. State of science: Pacemaker and defibrillator interference from wireless communication devices. Pacing Clin Electrophysiol. 1996;19:14191430[CrossRef][Medline]
- Santucci PA, Haw J, Trohman RG, Pinski SL. Interference with an implantable defibrillator by an electronic antitheft-surveillance device. N Engl J Med. 1998;339:13711374
[Free Full Text] - Groh WJ, Boschee SA, Engelstein ED. Interactions between electronic article surveillance systems and implantable cardioverter-defibrillators. Circulation. 1999;100:387392
[Abstract/Free Full Text] - Callans DJ, Hook BG, Kleiman RB, Mitra RL, Flores BT, Marchlinski FE. Unique sensing errors in third generation implantable cardioverter-defibrillators. J Am Coll Cardiol. 1993;22:11351140[Abstract]
- Fetter JG, Ivans V, Benditt DG, Collins J. Digital cellular telephone interaction with implantable cardioverter-defibrillators. J Am Coll Cardiol. 1998;31:623628
[Abstract/Free Full Text] - Occhetta E, Plebani L, Bortnik M, Sacchetti G, Trevi G. Implantable cardioverter defibrillators and cellular telephones: is there any interference? Pacing Clin Electrophysiol. 1999;22:983989[CrossRef][Medline]
- Sanmartin M, Fernandez LI, Marquez J. The absence of interference between GSM mobile telephones and implantable defibrillators: an in-vivo study. Rev Esp Cardiol. 1997;50:715719[ISI][Medline]
- Bassen HI, Moore HJ, Ruggera PS. Cellular phone interference testing of implantable cardiac defibrillators in vitro. Pacing Clin Electrophysiol. 1998;21:17091715[CrossRef][Medline]
- Barbaro V, Bartolini P, Bellocci F. Electromagnetic interference of digital and analog cellular telephones withimplantable cardioverter defibrillators: in vitro and in vivo studies. Pacing Clin Electrophysiol. 1999;22:626634[CrossRef][Medline]
- Irnich W. Implantable defibrillators and electromagnetic interference. Pacing Clin Electrophysiol. 1999;22:981982[CrossRef][Medline]
- Manolis AS, Chiladakis J, Vassilikos V, Maounis Th, Cokkinos DV. Pectoral cardioverter defibrillators: Comparison of prepectoral and submuscular implantation techniques. Pacing Clin Electrophysiol. 1999;22:469478[CrossRef][Medline]
- Haynes DL, Wang PJ, Reynolds DW. Interference with cardiac pacemakers by cellular telephones. N Engl J Med. 1997;336:14731479
[Abstract/Free Full Text] - Hayes DL, Garrillo RG, Findlay G, Embrey M. State of the science: Pacemaker and defibrillator interference from wireless communication devises. Pacing Clin Electrophysiol. 1996;19:14191430[CrossRef][Medline]
This article has been cited by other articles:
![]() |
C. Kolb, S. Schmieder, G.u. Lehmann, B. Zrenner, M. R. Karch, A. Plewan, and C. Schmitt Do airport metal detectors interfere with implantable pacemakers or cardioverter-defibrillators? J. Am. Coll. Cardiol., June 4, 2003; 41(11): 2054 - 2059. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Niehaus, R. Gille, R. Cierpka, T. Korte, and J. Tebbenjohans Interference of two common European digital cellular phones with implantable cardioverter-defibrillators Eur. Heart J., April 1, 2002; 23(7): 586 - 588. [Full Text] [PDF] |
||||
![]() |
A.S. Manolis and J.A. Chiladakis A reply Eur. Heart J., April 1, 2002; 23(7): 588 - 588. [Full Text] [PDF] |
||||
![]() |
M. Santini Digital cellular telephones and ICDs Eur. Heart J., August 1, 2001; 22(15): 1251 - 1252. [Abstract] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||