Cost-effectiveness of dual-chamber pacemaker therapy: does single lead VDD pacing reduce treatment costs of atrioventricular block?
a Medical University of Lübeck, Department of Internal Medicine II, Lübeck, Germany
b Department of Finances, Lübeck, Germany
revised January 10, 2000; accepted January 14, 2000
Abstract
Aims Implantation of single-lead VDD pacemakers is an established alternative to DDD pacing in patients with atrioventricular block. This study compares the long-term costs of both systems.
Methods and Results Three hundred and sixty patients with atrioventricular block received VDD or DDD pacemakers in alternating order. Primary costs of implantation included: devices, leads and operation material, surgeons, nurses, medical technicians, and hospitalization. The mean cost of an uncomplicated DDD pacemaker implantation was defined as 1000 virtual cost-units (CU). Costs of pacemaker related complications or re-operations as well as upgrades from VDD to DDD devices were considered secondary costs and assessed during a mean follow-up period of 42±15 months. Pacing efficacy was assessed by event-free survival with maintained atrioventricular synchronized pacing mode. Costs of pacemaker devices were not different (639±26CU in VDD vs 641±32CU in DDD, ns). However, due to lower costs of lead hardware (102±10CU in VDD vs 133±14CU in DDD, P<0·001) and shorter implantation procedures (44·3±5·1min vs 74·4±13·5min, P<0·001), costs of an uncomplicated implantation were 8·9% lower in the VDD group (911±35CU vs 1000±39CU,P <0·001). A smaller complication rate in the VDD group led to a 16·1% reduction of secondary costs (26±17CUyear1vs 31± 25CUyear1,P =0·024). Event-free survival did not differ between groups (83·4% in VDD vs 84·9% in DDD, ns).
Conclusion Use of single-lead VDD pacemakers achieves significant reduction of implantation and follow-up costs without loss of therapeutic efficacy compared to conventional DDD systems.
Key Words: Cost-effectiveness, dual-chamber pacing, single-lead VDD pacing, complications
f1 Correspondence: Dr med. Uwe K. H. Wiegand, Medizinische Universität zu Lübeck, Medizinische Klinik II, Ratzeburger Allee 16013538 Lübeck, Germany.
References
- Lamas GA, Orav EJ, Stambler BS. Quality of life and clinical outcomes in elderly patients treated with ventricular pacing as compared with dual-chamber pacing. Pacemaker Selection in the Elderly Investigators. N Engl J Med. 1998;338:10971104
[Abstract/Free Full Text] - Alpert MA, Curtis JJ, Sanfelippo JF. Comparative survival after permanent ventricular and dual chamber pacing for patients with chronic high degree atrioventricular block with and without preexistent congestive heart failure. J Am Coll Cardiol. 1986;7:925932[Abstract]
- Linde-Edelstamm C, Gullberg B, Nordlander R, Pehrsson SK, Rosenqvist M, Ryden L. Longevity in patients with high degree atrioventricular block paced in the atrial synchronous or the fixed rate ventricular inhibition mode. Pacing Clin Elektrophysiol. 1992;15:304313
- Nowak B, Voigtlander T, Himmrich E. Cardiac output in single-lead VDD pacing versus rate-matched VVIR pacing. Am J Cardiol. 1995;75:904907[CrossRef][ISI][Medline]
- Clarke M, Sutton R, Ward D, Camm A, Rickards A, Ingram A. Recommendations for pacemaker prescription for symptomatic patients. Report of a Working Party of the British Pacing and Elektrophysiology group. Br Heart J. 1991;66:185191
[Free Full Text] - Gregoratos G, Cheitlin MD, Conill A. ACC/AHA guidelines for implantation of cardiac pacemakers and antiarrhythmic devices: a report of the ACC/AHA Task Force on Practice Guidelines (Committee on Pacemaker Implantation). J Am Coll Cardiol. 1998;31:11751206
[Free Full Text] - De Belder MA, Linker NJ, Jones S, Camm AJ, Ward DE. Cost implications of the British Pacing and Electrophysiology Group's recommendations for pacing. Br Med J. 1992;305:861865[ISI][Medline]
- Ray SG, Griffith MJ, Jamieson S, Bexton RS, Gold RG. Impact of the recommendations of the British Pacing and Electrophysiology Group on pacemaker prescription and on the immediate costs of pacing in the Northern Region. Br Heart J. 1992;68:531534[Medline]
- Sutton R, Bourgeois I. Cost benefit analysis of single and dual chamber pacing for sick sinus syndrome and atrioventricular block. An economic sensitivity analysis of the literature. Eur Heart J. 1996;17:574582
[Abstract/Free Full Text] - Wiegand UKH, Bode F, Schneider R. Atrial sensing and AV-synchrony in single-lead VDD-pacemakers. A prospective comparison to DDD-devices with bipolar atrial leads. J Cardiovasc Elektrophysiol. 1999;10:513520
- Schriftenreihe des Bundesministerium für Gesundheit. 1995;45:
- Ansani L, Perocco GF, Guardigli G. Long-term reliability of single lead atrial synchronous pacing systems using closely spaced atrial dipoles: five year experience. Pacing Clin Electrophysiol. 1994;17:18651869[CrossRef][Medline]
- Crick J. European multicenter prospective follow-up study of 1002 implants of a single-lead VDD pacing system. Pacing Clin Elektrophysiol. 1991;14:18541857
- Curcio J. A multicenter evaluation of a single-pass lead VDD system. Pacing Clin Elektrophysiol. 1991;14:434442
- Folino AF, Buja G, Ruzza L, Nava A. Long-term follow-up of patients with single lead VDD stimulation. Pacing Clin Elektrophysiol. 1994;17:18541858[CrossRef]
- Ferguson TB Jr, Ferguson CL, Crites K, Crimmins-Reda P. The additional hospital costs generated in the management of complications of pacemaker and defibrillator implantations. J Thorac Cardiovasc Surg. 1996;111:742751
[Abstract/Free Full Text] - Antonelli D, Rosenfeld T, Freedberg NA, Palma E, Gross JN, Furman S. Insulation lead failure: is it a matter of insulation coating, venous approach, or both? Pacing Clin Electrophysiol. 1998;21:418421[CrossRef][Medline]
- Aggarwal RK, Connelly DT, Ray SG, Ball J, Charles RG. Early complications of permanent pacemaker implantation: no difference between dual and single chamber systems. Br Heart J. 1995;73:571575
[Abstract/Free Full Text] - Gross JN, Moser S, Benedek ZM, Andrews C, Furman S. DDD pacing mode survival in patients with a dual-chamber pacemaker. J Am Coll Cardiol. 1992;19:15361541[Abstract]
- Ibrahim B, Sanderson JE, Wright B, Palmer R. Dual chamber pacing: how many patients remain in DDD-mode over a long term? Br Heart J. 1995;74:7679
[Abstract/Free Full Text] - Sethi KK, Bhargava M, Mohan JC. Long term performance of atrial leads. Indian Heart J. 1994;46:7175[Medline]
- Andersen HR, Thuesen L, Bagger JP, Vesterlund T, Thomsen EB. Prospective randomised trial of atrial versus ventricular pacing in sick sinus syndrome. Lancet. 1994;344:15231528[CrossRef][ISI][Medline]
- Wiegand UKH, Bode F, Schneider R. Occurrence of sinus node disease in patients with AV-block. Implications for single-lead VDD-pacing. Heart. 1999;81:580585
[Abstract/Free Full Text] - Chamberlain-Webber R, Petersen ME, Ingram A, Briers L, Sutton R. Reasons for reprogramming dual chamber pacemakers to VVI mode: a retrospective review using a computer database. Pacing Clin Electrophysiol. 1994;17:17301736[CrossRef][Medline]
- Chamberlain-Webber R, Barnes E, Papouchado M, Crick JP. Long-term survival of VDD pacing. Pacing Clin Electrophysiol. 1998;21:22462248[CrossRef][Medline]
- Crossley GH, Gayle DD, Simmons TW. Reprogramming pacemakers enhances longevity and is cost-effective. Circulation. 1996;94:II-2457
- Die Gebührenordnung für Ärzte (GOÄ), 1996
This article has been cited by other articles:
![]() |
H. M. Spotnitz Surgical Implantation of Pacemakers and Automatic Defibrillators Card. Surg. Adult, January 1, 2008; 3(2008): 1395 - 1428. [Full Text] |
||||
![]() |
M. Kuhne, B. Schaer, C. Kaufmann, N. Moulay, T. Cron, T. Cueni, P. Weiss, C. Schindler, C. Sticherling, and S. Osswald A randomized trial comparing two different approaches of pacemaker selection Europace, December 1, 2007; 9(12): 1185 - 1190. [Abstract] [Full Text] [PDF] |
||||
![]() |
Authors/Task Force Members, P. E. Vardas, A. Auricchio, J.-J. Blanc, J.-C. Daubert, H. Drexler, H. Ector, M. Gasparini, C. Linde, F. B. Morgado, et al. Guidelines for cardiac pacing and cardiac resynchronization therapy: The Task Force for Cardiac Pacing and Cardiac Resynchronization Therapy of the European Society of Cardiology. Developed in Collaboration with the European Heart Rhythm Association Europace, October 1, 2007; 9(10): 959 - 998. [Full Text] [PDF] |
||||
![]() |
Authors/Task Force Members, P. E. Vardas, A. Auricchio, J.-J. Blanc, J.-C. Daubert, H. Drexler, H. Ector, M. Gasparini, C. Linde, F. B. Morgado, et al. Guidelines for cardiac pacing and cardiac resynchronization therapy: The Task Force for Cardiac Pacing and Cardiac Resynchronization Therapy of the European Society of Cardiology. Developed in Collaboration with the European Heart Rhythm Association Eur. Heart J., September 2, 2007; 28(18): 2256 - 2295. [Full Text] [PDF] |
||||
![]() |
I. Zupan, L. Lipar, D. Zizek, W. Boute, M. Vidmar, T. Gabrijelcic, P. Rakovec, and A. Brecelj Retrospective analysis of mode survival, reliability of atrial sensing and incidence of atrial tachyarrhythmias in 307 single-lead VDD pacemaker patients Europace, October 1, 2006; 8(10): 855 - 858. [Abstract] [Full Text] [PDF] |
||||
![]() |
F Eberhardt, F Bode, H Bonnemeier, F Boguschewski, M Schlei, W Peters, and U K H Wiegand Long term complications in single and dual chamber pacing are influenced by surgical experience and patient morbidity Heart, April 1, 2005; 91(4): 500 - 506. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Brunner, M. Olschewski, A. Geibel, C. Bode, and M. Zehender Long-term survival after pacemaker implantation: Prognostic importance of gender and baseline patient characteristics Eur. Heart J., January 1, 2004; 25(1): 88 - 95. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. M. Spotnitz Pacemakers and Automatic Defibrillators Card. Surg. Adult, January 1, 2003; 2(2003): 1293 - 1326. [Full Text] |
||||
![]() |
S. Pakarinen and L. Toivonen Pre-implant determinants of adequate long-term function of single lead VDD pacemakers Europace, January 1, 2002; 4(2): 137 - 141. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||