European Heart Journal Advance Access originally published online on February 25, 2005
European Heart Journal 2005 26(13):1276-1283; doi:10.1093/eurheartj/ehi161
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Randomized comparison of intra-aortic balloon support with a percutaneous left ventricular assist device in patients with revascularized acute myocardial infarction complicated by cardiogenic shock
Department of Internal Medicine/Cardiology, University of Leipzig-Heart Centre, Strümpellstr. 39, 04289 Leipzig, Germany
Received 22 July 2004; revised 25 November 2004; accepted 6 January 2005; online publish-ahead-of-print 25 February 2005.
* Corresponding author. Tel: +49 341 865 1428; fax: +49 341 865 1461. E-mail address: thielh{at}medizin.uni-leipzig.de
Aims Mortality in cardiogenic shock (CS) following acute myocardial infarction (AMI) remains unacceptably high despite percutaneous coronary intervention (PCI) of the infarcted artery and use of intra-aortic balloon pump (IABP) counterpulsation. A newly developed percutaneous left ventricular assist device (VAD) (Tandem HeartTM, Cardiac Assist, Pittsburgh, PA, USA) with active circulatory support might have positive haemodynamic effects and decrease mortality.
Methods and results Patients in CS after AMI, with intended PCI of the infarcted artery, were randomized to either IABP (n=20) or percutaneous VAD support (n=21). The primary outcome measure cardiac power index, as well as other haemodynamic and metabolic variables, could be improved more effectively by VAD support from 0.22 [interquartile range (IQR) 0.190.30] to 0.37 W/m2 (IQR 0.300.47, P<0.001) when compared with IABP from 0.22 (IQR 0.180.30) to 0.28 W/m2 (IQR 0.240.36, P=0.02; P=0.004 for intergroup comparison). However, complications like severe bleeding (n=19 vs. n=8, P=0.002) or limb ischaemia (n=7 vs. n=0, P=0.009) were encountered more frequently after VAD support, whereas 30 day mortality was similar (IABP 45% vs. VAD 43%, log-rank, P=0.86).
Conclusion Haemodynamic and metabolic parameters can be reversed more effectively by VAD than by standard treatment with IABP. However, more complications were encountered by the highly invasive procedure and by the extracorporeal support.
Key Words: Shock Heart-assist device Extracorporeal circulation Myocardial infarction
![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
J. M. Cheng, C. A. den Uil, S. E. Hoeks, M. van der Ent, L. S.D. Jewbali, R. T. van Domburg, and P. W. Serruys Percutaneous left ventricular assist devices vs. intra-aortic balloon pump counterpulsation for treatment of cardiogenic shock: a meta-analysis of controlled trials Eur. Heart J., September 1, 2009; 30(17): 2102 - 2108. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. D. Sjauw, A. E. Engstrom, M. M. Vis, R. J. van der Schaaf, J. Baan Jr, K. T. Koch, R. J. de Winter, J. J. Piek, J. G.P. Tijssen, and J. P.S. Henriques A systematic review and meta-analysis of intra-aortic balloon pump therapy in ST-elevation myocardial infarction: should we change the guidelines? Eur. Heart J., February 2, 2009; 30(4): 459 - 468. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Thiele and G. Schuler Cardiogenic shock: to pump or not to pump? Eur. Heart J., February 2, 2009; 30(4): 389 - 390. [Full Text] [PDF] |
||||
![]() |
H. Thiele, C. Kaulfersch, I. Daehnert, M. Schoenauer, I. Eitel, M. Borger, and G. Schuler Immediate primary transcatheter closure of postinfarction ventricular septal defects Eur. Heart J., January 1, 2009; 30(1): 81 - 88. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Seyfarth, D. Sibbing, I. Bauer, G. Frohlich, L. Bott-Flugel, R. Byrne, J. Dirschinger, A. Kastrati, and A. Schomig A Randomized Clinical Trial to Evaluate the Safety and Efficacy of a Percutaneous Left Ventricular Assist Device Versus Intra-Aortic Balloon Pumping for Treatment of Cardiogenic Shock Caused by Myocardial Infarction J. Am. Coll. Cardiol., November 4, 2008; 52(19): 1584 - 1588. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Takagaki, C. Wurzer, R. Wade, R. Lee, S. C. Malaisrie, P. M. McCarthy, and E. C. McGee Jr Successful Conversion of TandemHeart Left Ventricular Assist Device to Right Ventricular Assist Device After Implantation of a HeartMate XVE Ann. Thorac. Surg., November 1, 2008; 86(5): 1677 - 1679. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. D. Gregoric, G. Poglajen, M. Span, O.H. Frazier, P. Loyalka, and B. Kar Percutaneous Ventricular Assist Device Support During Off-Pump Surgical Coronary Revascularization Ann. Thorac. Surg., August 1, 2008; 86(2): 637 - 639. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. R. Reynolds and J. S. Hochman Cardiogenic Shock: Current Concepts and Improving Outcomes Circulation, February 5, 2008; 117(5): 686 - 697. [Full Text] [PDF] |
||||
![]() |
J. M. Wilson and J. T. Willerson Myocardial Revascularization with Percutaneous Devices Card. Surg. Adult, January 1, 2008; 3(2008): 573 - 598. [Full Text] |
||||
![]() |
M. Kooshkabadi, A. Kalogeropoulos, V. C. Babaliaros, and S. Lerakis Transesophageal guided left atrial positioning of a percutaneous ventricular assist device Eur J Echocardiogr, January 1, 2008; 9(1): 175 - 177. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. A. Pitsis, A. N. Visouli, D. Burkhoff, P. Dardas, N. Mezilis, G. Bougioukas, and G. Filippatos Feasibility Study of a Temporary Percutaneous Left Ventricular Assist Device in Cardiac Surgery Ann. Thorac. Surg., December 1, 2007; 84(6): 1993 - 1999. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. V. Jeger, A. M. Lowe, C. E. Buller, M. E. Pfisterer, V. Dzavik, J. G. Webb, J. S. Hochman, U. P. Jorde, and for the SHOCK Investigators Hemodynamic Parameters Are Prognostically Important in Cardiogenic Shock But Similar Following Early Revascularization or Initial Medical Stabilization: A Report From the SHOCK Trial Chest, December 1, 2007; 132(6): 1794 - 1803. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Thiele, R. W. Smalling, and G. C. Schuler Percutaneous left ventricular assist devices in acute myocardial infarction complicated by cardiogenic shock Eur. Heart J., September 1, 2007; 28(17): 2057 - 2063. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. G. Leshnower, T. G. Gleason, M. L. O'Hara, A. Pochettino, Y. J. Woo, R. J. Morris, T. J. Gardner, and M. A. Acker Safety and Efficacy of Left Ventricular Assist Device Support in Postmyocardial Infarction Cardiogenic Shock Ann. Thorac. Surg., April 1, 2006; 81(4): 1365 - 1371. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Mancini and D. Burkhoff Mechanical Device-Based Methods of Managing and Treating Heart Failure Circulation, July 19, 2005; 112(3): 438 - 448. [Abstract] [Full Text] [PDF] |
||||