European Heart Journal Advance Access published online on March 30, 2005
European Heart Journal, doi:10.1093/eurheartj/ehi163
1 Center for Cardiovascular Disease, University of Freiburg Medical Center, Hugstetterstrasse 55, 79106 Freiburg, Germany
* To whom correspondence should be addressed. Aims A lack of donor hearts has stimulated interest in using blood pumps to treat severe heart failure. We tested the hypothesis that a new continuous flow circulatory assist device could be employed safely to relieve symptoms of heart failure and evaluated the potential to prolong life. Methods and results An intracardiac axial flow pump was implanted in 17 heart failure patients [idiopathic dilated (12), ischaemic (4), or amyloid cardiomyopathy (1)]. All were deemed ineligible for transplantation. Implantation of the device was by left thoracotomy (15) or median sternotomy (2). Power delivery was by a skull-mounted titanium pedestal. All patients survived surgery. None needed right ventricular support. There were three hospital deaths, two early from subdural haematoma and aortic thrombosis, one late after switching to transplantation. A total of 14 patients left hospital with a cumulative support-time of 15.9 years (median: 293 days, interquartile range: 286 days, 1-44 months). Actuarial 1-, 2-, and 3-year survivals were 56, 47, and 24%, respectively. There was no pump failure. Quality of life scores improved. Two superficial pedestal infections were successfully treated. Four patients had cerebral thrombo-embolism: two early events attributed to inadequate anticoagulation and two late with near-complete resolution. An improved anticoagulant regime addressed this problem. Late death occurred in five patients from battery disconnection, subdural haematoma, bowel ischaemia, respiratory failure, and after cardiac transplantation. Conclusion Continuous flow blood pumps provided symptomatic relief of severe heart failure with high quality of life. Event-free survival reached 4 years. Analysis of adverse events led to improved management strategies. There is potential for widespread use of blood pumps in the community. A controlled trial is required.
Received August 9, 2004
Revised January 4, 2005
Accepted January 6, 2005
Clinical research
Advanced heart failure: feasibility study of long-term continuous axial flow pump support
2 John Radcliffe Hospital, Oxford, UK
3 Department of Cardiovascular Surgery, Texas Heart Institute, Houston, TX, USA
4 Royal Brompton and Harefield Hospital Trust and Imperial College, London, UK
Michael P. Siegenthaler, E-mail: siegenth{at}ch11.ukl.uni-freiburg.de
![]()
Abstract ![]()
CiteULike
Connotea
Del.icio.us What's this?
This article has been cited by other articles:
![]() |
C. Loffler, A. Straub, N. Bassler, K. Pernice, F. Beyersdorf, C. Bode, M. P. Siegenthaler, and K. Peter Evaluation of platelet activation in patients supported by the Jarvik 2000* high-rotational speed impeller ventricular assist device. J. Thorac. Cardiovasc. Surg., March 1, 2009; 137(3): 736 - 741. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Geisen, C. Heilmann, F. Beyersdorf, C. Benk, M. Berchtold-Herz, C. Schlensak, U. Budde, and B. Zieger Non-surgical bleeding in patients with ventricular assist devices could be explained by acquired von Willebrand disease Eur. J. Cardiothorac. Surg., April 1, 2008; 33(4): 679 - 684. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Markl, C. Benk, D. Klausmann, A. F. Stalder, A. Frydrychowicz, J. Hennig, and F. Beyersdorf Three-dimensional magnetic resonance flow analysis in a ventricular assist device. J. Thorac. Cardiovasc. Surg., December 1, 2007; 134(6): 1471 - 1476. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. B. Selvanayagam, P. N. Hawkins, B. Paul, S. G. Myerson, and S. Neubauer Evaluation and Management of the Cardiac Amyloidosis J. Am. Coll. Cardiol., November 27, 2007; 50(22): 2101 - 2110. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. P. Siegenthaler, O.H. Frazier, F. Beyersdorf, J. Martin, H. Laks, J. Elefteriades, A. Khaghani, U. Kjellman, B. Koul, J. Pepper, et al. Mechanical reliability of the jarvik 2000 heart. Ann. Thorac. Surg., May 1, 2006; 81(5): 1752 - 1759. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. C. Deng The challenges of generating evidence to guide mechanical circulatory support-based management of advanced heart failure Eur. Heart J., May 2, 2005; 26(10): 953 - 955. [Full Text] [PDF] |
||||