Impact of cardiac magnetic resonance imaging on human lymphocyte DNA integrity

(CC)
Michael Fiechter, Julia Stehli, Tobias A. Fuchs, Svetlana Dougoud, Oliver Gaemperli, Philipp A. Kaufmann
DOI: http://dx.doi.org/10.1093/eurheartj/eht184 2340-2345 First published online: 22 June 2013

Abstract

Aims Magnetic resonance (MR) imaging is widely used for diagnostic imaging in medicine as it is considered a safe alternative to ionizing radiation-based techniques. Recent reports on potential genotoxic effects of strong and fast switching electromagnetic gradients such as used in cardiac MR (CMR) have raised safety concerns. The aim of this study was to analyse DNA double-strand breaks (DSBs) in human blood lymphocytes before and after CMR examination.

Methods and results In 20 prospectively enrolled patients, peripheral venous blood was drawn before and after 1.5 T CMR scanning. After density gradient cell separation of blood samples, DNA DSBs in lymphocytes were quantified using immunofluorescence microscopy and flow cytometric analysis. Wilcoxon signed-rank testing was used for statistical analysis. Immunofluorescence microscopic and flow cytometric analysis revealed a significant increase in median numbers of DNA DSBs in lymphocytes induced by routine 1.5 T CMR examination.

Conclusion The present findings indicate that CMR should be used with caution and that similar restrictions may apply as for X-ray-based and nuclear imaging techniques in order to avoid unnecessary damage of DNA integrity with potential carcinogenic effect.

  • Cardiac MRI
  • DNA damage
  • γ-H2AX
  • Flow cytometry
  • Immunofluorescence microscopy

See page 2337 for the editorial comment on this article (doi:10.1093/eurheartj/eht214)

Introduction

Magnetic resonance (MR) imaging is a widely used and well-established non-invasive medical diagnostic imaging tool. By using a static and a gradient magnetic field in combination with a radiofrequency field (RF), MR provides excellent contrast among different tissues of the body including the brain, musculoskeletal system, and the heart. Although long-term effects on human health from exposure to strong static magnetic fields seem unlikely,1 acute effects such as vertigo, nausea, change in blood pressure, reversible arrhythmia,2 and neurobehavioural effects have been documented from occupational exposition to 1.5 T.3 Cardiac MR (CMR) imaging requires some of the strongest and fastest switching electromagnetic gradients available in MR exposing the patients to the highest administered energy levels accepted by the controlling authorities.4 Studies focusing on experimental teratogenic59 or carcinogenic1012 effects of MR revealed conflicting results. Since CMR is emerging as one of the fastest growing new fields of broad MR application,13 it is of particular concern that a recent in vitro study with CMR sequences has reported on CMR-induced DNA damages in white blood cells up to 24 h after exposure to 1.5 T CMR.4 It is in this context that the European Parliament,14 the International Commission on Non-Ionizing Radiation Protection (ICNIRP),15,16 and the World Health Organization (WHO)17 have urgently called for an action in order to evaluate adverse biological effects of clinical MR scanning.

The aim of the present study was to assess the impact of routine CMR scanning on DNA double-strand breaks (DSBs) of peripheral blood mononuclear cells (PBMCs) as a measure of the carcinogenic potential of this examination.

Methods

Twenty consecutive patients referred for cardiac evaluation were included. After obtaining written informed consent, 10 mL of peripheral blood was drawn before and after undergoing routine contrast (gadobutrolum, Gadovist, Bayer Schering Pharma, Germany) enhanced CMR examination18 on a 1.5 T MR scanner (Philips Achieva, Best, NL, USA) as approved by the local ethics committee (KEK-Nr. 849). PBMCs were obtained using density gradient separation (Histopaque 1077, Sigma-Aldrich) as previously established.19

The clinical CMR protocol used in our daily routine has been recently reported in detail.20 In brief, a commercially available MR scanner (Philips 1.5 T, Achieva, software release 3.2.1) equipped with a maximum gradient strength of 42 mT/m and a maximum gradient speed of 180 mT/m/ms was used. The following standard pulse sequences to generate images were used: gradient echo, steady-state free precession, FastSE, T2-weighted double-inversion black-blood spin-echo sequence for oedema imaging, balanced SSFP sequence for perfusion and inversion recovery segmented gradient echo sequence for late gadolinium enhancement.

DSBs were detected by immunofluorescence microscopy using a rabbit-anti-human phospho-histone γ-H2AX and a goat-anti-rabbit-AlexaFluor-488 antibody (CST Cell Signalling Technology, adapted from May et al.21). Cell nuclei were counterstained with 4′,6-diamidino-2-phenylindole (DAPI, Vector Laboratories) and the γ-H2AX foci per lymphocyte were visualized on an inverse confocal microscope (CLSM-Model SP5, Leica Microsystems) and quantified by a blinded observer.

With flow cytometry (FACScanto, BD Bioscience), DSBs were additionally quantified in T-lymphocytes22,23 previously identified by a mouse-anti-human CD3-APC antibody (Life Technologies). Based on forward and side light scattering, PBMCs were gated for viable single-cell events and proper compensation controls were used in flow cytometric analyses to correct for spectral overlap. Data from flow cytometric quantification (MFI, geometric mean of fluorescence intensity of γ-H2AXpositive T-lymphocytes) was evaluated using FlowJo software (V10.0.2, Tree Star, Inc.).

Based on a variation of γ-H2AX assessment at 20% as reported by Muslimovic et al.,22 an average difference in γ-H2AX findings reported in ex vivo experiments,4 aiming at alpha = 0.05 and a power (1 − β) of 0.8, the number of patients necessary was calculated between 10 and 15.

SPSS 20.0 (SPSS, Chicago, IL, USA) was used for all statistical analysis. The Shapiro–Wilk test was applied to exclude normal distribution of data sets. This was followed by testing for significant differences between DSBs before and after CMR examination by using the Wilcoxon signed-rank test. P-values of <0.05 (two-tailed) were considered statistically significant.

Results

Mean age of patients was 53 ± 13 years and 16 (80%) were males. Ten patients were referred for evaluation of cardiomyopathy and 10 for the assessment of myocardial ischaemia. The mean CMR scan duration was 68 ± 22 min with an average contrast media bolus of 15 ± 4 mL. The patient baseline characteristics are given in Table 1.

View this table:
Table 1

Patient baseline characteristics (n = 20)

Age (years ± SD)53 ± 13
BMI (kg/m2 ± SD)25 ± 4
Male, n (%)16 (80)
Cardiovascular risk factors, n (%)
 Arterial hypertension6 (30)
 Diabetes mellitus4 (20)
 Dyslipidaemia4 (20)
 Smoking2 (10)
 Positive family history1 (5)
Medications, n (%)
 Aspirin7 (35)
 Beta-blocker9 (45)
 ACE/angiotensin II inhibitor8 (40)
 Statin7 (35)
  • SD, standard deviation; BMI, body mass index.

By immunofluorescence microscopy (Figure 1), the median number of DSBs (foci, Table 2) per lymphocyte in baseline samples was 0.066 (range: 0–0.661) and increased significantly (P < 0.05) after CMR exposure to 0.190 (range: 0–1.065, Figure 2).

View this table:
Table 2

Increase in double-strand breaks after cardiac magnetic resonance assessed by immunofluorescence

 Microscopy foci per lymphocyteFlow cytometry MFI
BeforeAfterBeforeAfter
Mean0.1430.270*29893395*
SD0.1910.227850906
Median0.0660.190*27583232*
MAD0.1370.199640696
IQR0.1690.25711331198
  • IF, immunofluorescence (units are foci per lymphocyte); MFI, geometric mean of T-lymphocyte fluorescence intensity (arbitrary units); γ-H2AX, marker of DSBs; SD, standard deviation; MAD, median absolute deviation; IQR, interquartile range.

  • *Indicates P < 0.05 vs. before.

Figure 1

Visualization of double-strand breaks (DSBs) in nuclei (arrow heads) of human lymphocytes of two patients before and after cardiac magnetic resonance scans by immunofluorescence microscopy. DSBs (foci, white arrows) are detected by γ-H2AX staining (green).

Figure 2

Amount of double-strand breaks before and after cardiac magnetic resonance (CMR) scan by immunofluorescence microscopy. After CMR scanning, there was a significant increase (*P < 0.05) in γ-H2AX foci per lymphocyte by immunofluorescence microscopy. Bars indicate median values with median absolute deviation (left panel) and individual values are interconnected with a line (right panel).

In T-lymphocytes, flow cytometry (Figure 3) revealed a median MFI (arbitrary units) of 2758 (range: 1907–5109) before and 3232 (range: 2413–5484) after CMR (P < 0.005, Table 2 and Figure 4).

Figure 3

Flow cytometric analysis of double-strand breaks (γ-H2AXpositive T-lymphocytes) before and after cardiac magnetic resonance (CMR) scan. T-lymphocytes were readily identified by representative dot plots and histograms (lymphocytes, DAPI, and CD3). The shift of the left curve (red, before CMR) to the right curve (blue, after CMR) in the presented overlay indicates an increase in double-strand breaks (γ-H2AXpositive T-lymphocytes). SSC-A: side scatter channel area. FSC-A: forward scatter channel area. DAPI: 4′,6-diamidino-2-phenylindole, counterstaining cell nuclei. CD3: mouse-anti-human CD3-APC antibody counterstaining specifically the T-lymphocytes.

Figure 4

Amount of double-strand breaks before and after cardiac magnetic resonance scan by flow cytometry of γ-H2AXpositive T-lymphocytes using geometric mean fluorescence intensity (MFI). The median MFI increased significantly after cardiac magnetic resonance scanning (*P < 0.005, left panel). Individual values are interconnected with a line (right panel).

Discussion

We show here that clinical routine CMR scanning exerts genotoxic effects. Although many experimental in vitro studies have suggested DNA damage after exposure to MR imaging, we present the first in vivo results documenting that contrast CMR scanning in daily clinical routine is associated with increased lymphocyte DNA damage.

The different components of the magnetic field during CMR may have contributed to the observed DNA damage. The gradient field generated during MR scanning includes extremely low frequencies (ELF), which have been classified by the International Agency for Research on Cancer (IARC) as possible human carcinogen (group 2B)24 based on a large body of literature on the genotoxic effects of ELF magnetic fields.2528 The latter seem to be involved directly and indirectly in DNA and chromosomal damage by inducing reactive oxygen species.29 Similarly, DNA damage and chromosome alterations have been discussed after exposure to RF.

Our results do not allow commenting on the persistence of the induced DNA damage, although this is a key issue of genetic risk assessment, because damage can trigger DNA instability and exert tumourigenic effects. Due to the long time delay between DSB induction and resulting cancer development, our study cannot quantify such long-term effects as this was beyond the scope of the present study. This, however, is true in principle for any observation of DSB induction from any diagnostic radiation exposure including ionizing radiation, for which no direct observational proof of its adverse impact on outcome is available due to the small scale of damage and the long delay between exposure and event. In view of the growing use of new generation MR scanners with increasing magnetic field strength (higher Tesla), our results seem to support the suggestions of the ICNIRP for an urgent need of monitoring workers and for epidemiologic studies on subjects with high levels of exposure or particular conditions such as for example pregnant occupational workers.30

Despite activation of repair mechanisms, persistence of DNA damage has been found in human lymphocytes more than 24 h after exposing patients and blood samples to CMR scanning.4 Co-genotoxic effects of MR in combination with the administered gadolinium-based contrast material may further have contributed to DNA damage due to the potentiating effect of gadolinium-based contrast material and MR exposure.31 As in our study all patients underwent contrast enhanced CMR, reflecting widely used clinical practice,32 we cannot differentiate the precise contribution of the known genotoxic effect of the gadolinium-based contrast material from the effects of the magnetic field. However, the use of contrast material is generally an integrated part of CMR scanning and therefore our results may appropriately represent the effect of a routine CMR scan. The absolute amount of DNA damage is certainly larger in our study compared with previous in vitro studies, as the entire blood of each patient rather than a blood sample was exposed during CMR. According to the assumptions used in the field of radiation protection, an increased number of DNA damages confer a linearly increased risk of cancer. Conversely, even a low number of DSBs may represent a carcinogenic risk according to the linear-no threshold theory. Our results compare well to the more than two-fold increase in DSBs induced by CMR and assessed by immunofluorescence microscopy as reported by Simi et al.,4 which was substantially less pronounced than the almost six-fold increase observed after cardiac CT by Kuefner et al.33 Although only a few data are available using FACS analyses for this low scale of signal, the excellent agreement between microscopy and FACS over a large range of signal including the present study strengthens the validity of our results.34

Of note, observations in several subsets of patients seem to suggest increased sensibilities to MRI exposition, as higher susceptibility for DNA damage by MRI has been found for example in lymphocytes of patients with Turner's syndrome.35 Thus, inappropriate examinations should be avoided and CMR should be used with caution and similar restrictions may apply as for X-ray-based and nuclear imaging techniques where the potential harm is carefully weighted against the obvious benefit offered by each examination in order to avoid unnecessary damage of DNA integrity with potential carcinogenic effect.

Funding

Grants from the Swiss National Science Foundation to P.A.K. and to M.F. are gratefully acknowledged.

Acknowledgements

We thank Christiane Koenig, Jose Maria Mateos, PhD, Stefano Ferrari, PhD, and Florian Mair, MSc, from the University of Zurich, Zurich, Switzerland, and Arnold von Eckardstein, MD, Institute for Clinical Chemistry, University Hospital Zurich, Zurich, Switzerland, for their invaluable advice and excellent technical support for the immunofluorescence analysis.

Conflict of interest: none declared.

Footnotes

  • Contributed equally to this work.

  • Guest edited by Jeroen Bax, Professor of Cardiology, Leiden University Medical Centre, Leiden, Netherlands.

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/), which permits non-commercial re-use, distribution, and reproduction in any medium, provided the original work is properly cited. For commercial re-use, please contact journals.permissions@oup.com.